A TRANSIENTLY HEATED NEEDLE ANEMOMETER

被引:10
作者
BLAND, WL
NORMAN, JM
CAMPBELL, GS
CALISSENDORFF, C
MILLER, EE
机构
[1] WASHINGTON STATE UNIV,DEPT AGRON & SOILS,PULLMAN,WA 99163
[2] SOILTRON,BURLINGTON,WA 98233
关键词
D O I
10.1016/0168-1923(94)02190-U
中图分类号
S3 [农学(农艺学)];
学科分类号
0901 ;
摘要
Convective heat transfer from cylinders warmed above ambient temperature is a standard technique for measuring windspeed. Available devices are typically fragile, expensive, and complex. We developed a heat-transfer anemometer that is robust, inexpensive, and simple. Referred to as a needle anemometer, the sensing portion of the device is a 20 mm length of tubing of 0.56 mm diameter, into which a heater and thermocouple are inserted. Needle temperature measured during alternating 5 s heating and non-heating periods allows estimation of the equilibrium temperature rise above ambient temperature for long heating times. This is inversely related to (windspeed)1/2 from 0.05 m s-1 to greater than 14 m s-1. In a field trial, the needle anemometer yielded windspeeds of +/- 0.2 m s-1 (S(y,x)) of those from a cup anemometer. Precipitation striking the needle caused overestimates of windspeed by up to 20%.
引用
收藏
页码:227 / 235
页数:9
相关论文
共 2 条
[1]   A NOTE ON HEAT TRANSPORT [J].
KANEMASU, ET ;
TANNER, CB .
BIOSCIENCE, 1968, 18 (04) :327-&
[2]  
MONTEITH JL, 1990, PRINCIPLES ENV PHYSI