PERFORMANCE OF A MECHANICAL HEAT SWITCH AT LOW TEMPERATURES

被引:16
作者
COLWELL, JH
机构
关键词
D O I
10.1063/1.1684194
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A mechanical heat switch used in calorimetry experiments down to 0.3 K is described and its performance analyzed in sufficient detail to be useful for design purposes. The thermal conductance of the indium-copper switch contacts was determined as a function of temperature and applied force over the range 0.3-2 K. These results are compared with other conductance measurements to evaluate choices of metals for heat switch contacts. © 1969 The American Institute of Physics.
引用
收藏
页码:1182 / &
相关论文
共 22 条