MEASUREMENT OF FORMALDEHYDE CONCENTRATIONS IN A SUB-ATMOSPHERIC STEAM-FORMALDEHYDE AUTOCLAVE

被引:7
作者
MARCOS, D [1 ]
WISEMAN, D [1 ]
机构
[1] UNIV BRADFORD,POSTGRAD SCH STUDIES PHARM,BRADFORD BD7 1DP,W YORKSHIRE,ENGLAND
关键词
D O I
10.1136/jcp.32.6.567
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
A method has been developed for measuring formaldehyde concentrations in a sub-atmospheric steam-formaldehyde autoclave. Data obtained using this method indicate that the concentration of formaldehyde in the chamber atmosphere is not homogeneous and that it decreases rapidly with time. The penetration of formaldehyde vapour into narrow tubes has also been investigated and was shown to be dependent on the length-to-bore ratio of the tubes. The formaldehyde concentration within the tubes could be increased by using a lower vacuum in the air removal stage at the beginning of the cycle.
引用
收藏
页码:567 / 575
页数:9
相关论文
共 13 条
[1]  
ALDER VG, 1968, STERILIZATION TECHNI, P141
[2]   DETERMINATION OF FORMALDEHYDE IN GAS MIXTURES BY CHROMOTROPIC ACID METHOD [J].
ALTSHULLER, A ;
MILLER, DL ;
SLEVA, SF .
ANALYTICAL CHEMISTRY, 1961, 33 (04) :621-+
[3]  
AYLIFFE GAJ, 1977, LANCET, V1, P256
[4]  
Carruthers J. E., 1936, T FARADAY SOC, V32, P195
[5]  
Ernst R, 1970, Bull Parenter Drug Assoc, V24, P119
[6]   STERILIZATION WITH GASEOUS ETHYLENE OXIDE - A REVIEW OF CHEMICAL AND PHYSICAL FACTORS [J].
ERNST, RR ;
DOYLE, JE .
BIOTECHNOLOGY AND BIOENGINEERING, 1968, 10 (01) :1-&
[7]  
ERNST RR, 1975, ACTA PHARM SUEC, V12, P44
[8]   PROCESSING URINARY ENDOSCOPES IN A LOW-TEMPERATURE STEAM AND FORMALDEHYDE AUTOCLAVE [J].
GIBSON, GL .
JOURNAL OF CLINICAL PATHOLOGY, 1977, 30 (03) :269-274
[9]  
HANDLOS V, 1977, ARCH PHARM CHEMI SCI, V5, P163
[10]   USE OF GASEOUS ANTIMICROBIAL AGENTS WITH SPECIAL REFERENCE TO ETHYLENE OXIDE [J].
KELSEY, JC .
JOURNAL OF APPLIED BACTERIOLOGY, 1967, 30 (01) :92-&