HEARING WITH THE BONE-ANCHORED HEARING-AID (BAHA, HC-200) COMPARED TO A CONVENTIONAL BONE-CONDUCTION HEARING-AID

被引:41
作者
CREMERS, CWRJ [1 ]
SNIK, AFM [1 ]
BEYNON, AJ [1 ]
机构
[1] UNIV HOSP NIJMEGEN,INST OTORHINOLARYNGOL,DEPT EAR NOSE & THROAT,POB 9101,6500 HB NIJMEGEN,NETHERLANDS
来源
CLINICAL OTOLARYNGOLOGY | 1992年 / 17卷 / 03期
关键词
BONE-ANCHORED HEARING AID; TITANIUM IMPLANT;
D O I
10.1111/j.1365-2273.1992.tb01842.x
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Sixteen patients have been fitted with a standard bone-anchored hearing aid (HC 200), to replace their conventional bone-conduction aid. The average pure tone threshold at 0.5, 1 and 2 kHz varied from 35 to 75 dB HL, with a sensorineural component varying from 0 to 30 dB HL. The patients' performance with the bone-anchored aid was compared to that with the conventional bone-conduction aid in an acoustic-free field. The maximum phoneme score in quiet was 100% in most patients; in 6 patients, the score with the bone-anchored aid was better (range from 5 to 10%). The speech-in-noise ratio was significantly better in 11 patients (range from - 1.4 to - 8 dB). None of the patients had poorer results on either test with the bone-anchored aid. The improved speech recognition was ascribed to better performance of the hearing aid in the higher frequency range (above 2 kHz) and to relatively less distortion.
引用
收藏
页码:275 / 279
页数:5
相关论文
共 8 条
[1]   A SPEECH-TO-NOISE RATIO TEST WITH THE BONE-ANCHORED HEARING-AID - A COMPARATIVE-STUDY [J].
CARLSSON, P ;
HAKANSSON, B ;
ROSENHALL, U ;
TJELLSTROM, A .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1986, 94 (04) :421-426
[2]   SPEECH AUDIOMETRY - EARPHONE AND SOUND FIELD [J].
DIRKS, DD ;
STREAM, RW ;
WILSON, RH .
JOURNAL OF SPEECH AND HEARING DISORDERS, 1972, 37 (02) :162-176
[3]   HEARING THRESHOLDS WITH DIRECT BONE CONDUCTION VERSUS CONVENTIONAL BONE CONDUCTION [J].
HAKANSSON, B ;
TJELLSTROM, A ;
ROSENHALL, U .
SCANDINAVIAN AUDIOLOGY, 1984, 13 (01) :3-13
[4]   THE BONE-ANCHORED HEARING-AID - PRINCIPAL DESIGN AND A PSYCHOACOUSTICAL EVALUATION [J].
HAKANSSON, B ;
TJELLSTROM, A ;
ROSENHALL, U ;
CARLSSON, P .
ACTA OTO-LARYNGOLOGICA, 1985, 100 (3-4) :229-239
[5]  
HAKANSSON B, 1990, ANN OTO RHINOL LARYN, V99, P1
[6]   SUGGESTED THRESHOLD SOUND PRESSURE LEVELS FOR FREQUENCY-MODULATED (WARBLE) TONES IN THE SOUND FIELD [J].
MORGAN, DE ;
DIRKS, DD ;
BOWER, DR .
JOURNAL OF SPEECH AND HEARING DISORDERS, 1979, 44 (01) :37-54
[7]  
PLOMP R, 1979, AUDIOLOGY, V18, P43
[8]   5-YEAR EXPERIENCE WITH SKIN-PENETRATING BONE-ANCHORED IMPLANTS IN THE TEMPORAL BONE [J].
TJELLSTROM, A ;
ROSENHALL, U ;
LINDSTROM, J ;
HALLEN, O ;
ALBREKTSSON, T ;
BRANEMARK, PI .
ACTA OTO-LARYNGOLOGICA, 1983, 95 (5-6) :568-575