SUPRATHRESHOLD STATIC PERIMETRY - INITIAL CLINICAL-TRIALS WITH THE FIELDMASTER AUTOMATED PERIMETER

被引:48
作者
KELTNER, JL
JOHNSON, CA
BALESTRERY, FG
机构
[1] UNIV CALIF DAVIS,SACRAMENTO MED CTR,DEPT NEUROL,SACRAMENTO,CA 95817
[2] UNIV CALIF DAVIS,SACRAMENTO MED CTR,DEPT NEUROL SURG,SACRAMENTO,CA 95817
关键词
D O I
10.1001/archopht.1979.01020010112003
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We describe our initial clinical trials comparing the Fieldmaster automated visual field plotter (Synemed, Inc) with Goldmann kinetic perimetry performed by highly skilled perimetrists. The Fieldmaster utilizes a quick, randomized test procedure referred to as “suprathreshold static perimetry.” The calibration and stimulus control features of the Fieldmaster have made it possible to derive approximate Goldmann-equivalent stimulus values and provide accurate, reproducible test conditions. Evaluation of 105 eyes (61 with defects, 44 normal) revealed a 98.5% detection rate and a 4% false alarm rate for the Fieldmaster in comparison with Goldmann kinetic perimetry. Based on these results, the Fieldmaster appears to have great potential value as a quantitative visual field screening device for the detection of visual field defects. © 1979, American Medical Association. All rights reserved.
引用
收藏
页码:260 / 272
页数:13
相关论文
共 19 条
[1]   OCULAR PRESSURE AND VISUAL FIELDS - A TEN-YEAR FOLLOW-UP STUDY [J].
ARMALY, MF .
ARCHIVES OF OPHTHALMOLOGY, 1969, 81 (01) :25-&
[2]  
AULHORN E, 1977, 2ND INT VIS FIELD S, V14, P17
[3]   SCOTOPTER - A HAND PROJECTION CAMPIMETER [J].
BENTOVIM, N .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1967, 64 (04) :780-&
[4]   AUTOMATIC DEVICE FOR RAPID ASSESSMENT OF CENTRAL VISUAL FIELD [J].
BUCHANAN, WM ;
GLOSTER, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1965, 49 (02) :57-&
[5]  
CASSIDY V, 1959, ARCH OPHTHALMOL-CHIC, V61, P589
[6]   AUTOMATION OF PERIMETRY [J].
FANKHAUSER, F ;
ROULIER, A ;
KOCH, P .
ALBRECHT VON GRAEFES ARCHIV FUR KLINISCHE UND EXPERIMENTELLE OPHTHALMOLOGIE, 1972, 184 (02) :126-+
[7]  
FANKHAUSER F, 1977, 2ND INT VIS FIELD S, V14, P7
[9]   FLASH PERIMETRY [J].
GLOSTER, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1970, 54 (10) :649-&
[10]  
Greve E. L., 1973, SINGLE MULTIPLE STIM