DRUG DELIVERY FROM HOLDING CHAMBERS WITH ATTACHED FACEMASK

被引:118
作者
EVERARD, ML
CLARK, AR
MILNER, AD
机构
[1] FISONS PLC, RES & DEV, RESP PHYS INHALAT DEV, LOUGHBOROUGH, ENGLAND
[2] ST THOMAS HOSP, DEPT PAEDIAT, LONDON SE1 7EH, ENGLAND
关键词
D O I
10.1136/adc.67.5.580
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
There is much interest in the use of holding chambers with an attached facemask to deliver aerosols from metered dose inhalers to infants. In order to study the influence of various design factors on the dose inhaled at different tidal volumes, a model was constructed in which a Starling ventilator was used to generate an inspiratory/expiratory cycle across a filter. Sodium cromoglycate was administered via a Nebuhaler and mask, Aerochamber and mask, and a coffee cup using tidal volumes of 25, 50, and 150 ml and the dose deposited upon the filter after six breaths was assayed using an ultraviolet spectrophotometric method. At the lowest tidal volume the high aerosol concentration in the smaller chamber enhanced drug delivery while at the highest tidal volume delivery was greatest from the larger chamber reflecting the larger dose available. Multiple breaths ensured that the dose inhaled per kilogram from each chamber was relatively large and also permitted significant drug delivery despite the introduction of a relatively large dead space between valve and filter. The dose delivered was increased by increasing the dose introduced into the chamber though not proportionately. These devices appear likely to deliver significant quantities of aerosol to infants, though drug delivery may be enhanced by the use of an appropriate valve.
引用
收藏
页码:580 / 585
页数:6
相关论文
共 20 条
  • [1] BEASLEY CRW, 1985, NJ MED J, V87, P854
  • [2] PEP-spacer: an adaptation for administration of MDI to infants
    BISGAARD, H
    OHLSSON, S
    [J]. ALLERGY, 1989, 44 (05) : 363 - 364
  • [3] INHALED BUDESONIDE FOR TREATMENT OF RECURRENT WHEEZING IN EARLY-CHILDHOOD
    BISGAARD, H
    MUNCK, SL
    NIELSEN, JP
    PETERSEN, W
    OHLSSON, SV
    [J]. LANCET, 1990, 336 (8716) : 649 - 651
  • [4] RELIABLE SALBUTAMOL ADMINISTRATION IN 6-MONTH-OLD TO 36-MONTH-OLD CHILDREN BY MEANS OF A METERED DOSE INHALER AND AEROCHAMBER WITH MASK
    CONNER, WT
    DOLOVICH, MB
    FRAME, RA
    NEWHOUSE, MT
    [J]. PEDIATRIC PULMONOLOGY, 1989, 6 (04) : 263 - 267
  • [5] POTENTIAL LIMITATIONS OF A CONICAL SPACER DEVICE IN SEVERE ASTHMA
    COX, ID
    WALLIS, PJW
    APPS, MCP
    [J]. BRITISH MEDICAL JOURNAL, 1984, 288 (6423) : 1044 - 1044
  • [6] COMPARISON OF 3 TECHNIQUES OF INHALATION ON THE AIRWAY RESPONSE TO TERBUTALINE
    CUSHLEY, MJ
    LEWIS, RA
    TATTERSFIELD, AE
    [J]. THORAX, 1983, 38 (12) : 908 - 913
  • [7] CLINICAL-EVALUATION OF A SIMPLE DEMAND INHALATION MDI AEROSOL DELIVERY DEVICE
    DOLOVICH, M
    RUFFIN, R
    CORR, D
    NEWHOUSE, MT
    [J]. CHEST, 1983, 84 (01) : 36 - 41
  • [8] DRUG DELIVERY FROM JET NEBULIZERS
    EVERARD, ML
    CLARK, AR
    MILNER, AD
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1992, 67 (05) : 586 - 591
  • [9] FREIGANG B, 1977, CAN MED ASSOC J, V117, P1308
  • [10] NEBUHALER TECHNIQUE
    GLEESON, JGA
    PRICE, JF
    [J]. BRITISH JOURNAL OF DISEASES OF THE CHEST, 1988, 82 (02): : 172 - 174