INSTRUMENTATION AND DOSIMETER-SIZE ARTIFACTS IN QUANTITATIVE THERMOLUMINESCENCE DOSIMETRY OF LOW-DOSE FIELDS

被引:55
作者
MEIGOONI, AS
MISHRA, V
PANTH, H
WILLIAMSON, J
机构
[1] WASHINGTON UNIV, SCH MED, EDWARD MALLINCKRODT INST RADIOL, ST LOUIS, MO 63110 USA
[2] BEST MED INT, SPRINGFIELD, VA 22153 USA
关键词
TLD; THERMOLUMINESCENCE; DOSIMETRY; BRACHYTHERAPY;
D O I
10.1118/1.597555
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Thermoluminescence dosimetry is extensively used for quantitative dose measurements in various irradiation fields such as dosimetry of brachytherapy sources. In this application, small doses on the order of 0.5 cGy must be accurately measured, which requires careful control of instrumentation, energy-dependence, and nonlinearity of detector response. Several investigators have observed the presence of some undesirable signals when the thermoluminescent dosimeters (TLDs) were read without any nitrogen gas flow in the TLD reader. Others have indicated that the “prereadout” annealing technique is the same as the “preirradiation” technique for doses above 10 cGy, but they have not extended their study to lower doses. The goal of this study is to investigate dependence of sensitivity and linearity of the TLD response to the flow of nitrogen gas in the TLD reader at low dose level, annealing technique, and TLD size. The effect of nitrogen flow sensitivity and linearity of two different sizes of lithium fluoride TLD-100 chips has been studied. Our data indicate a large standard deviation of TLD sensitivity, up to a factor of 2, when TLDs were read without nitrogen gas flow in the TLD reader. In addition, a large deviation from linearity was observed for doses below 5 cGy. When the reading-chamber was purged with nitrogen gas, dispersion of the responses of the TLDs that were exposed to the same dose fell to within 5%. At precision levels of 2% and 5%, the low dose limits are 1 cGy and 0.5 cGy, respectively, for large chips and 15 cGy and 1 cGy for small chips, if TLDs are read with nitrogen gas flow in the TLD reader. The full results of our investigation are presented. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:555 / 561
页数:7
相关论文
共 27 条
[1]  
AITKEN MJ, 1967, P INT C LUMINESCENCE, V8
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]   LITHIUM-FLUORIDE GLOW-PEAK GROWTH DUE TO ANNEALING [J].
BOOTH, LF ;
ATTIX, FH ;
JOHNSON, TL .
HEALTH PHYSICS, 1972, 23 (02) :137-+
[4]  
CAMERON J. R., 1964, HLTH PHYS, V10, P25, DOI 10.1097/00004032-196401000-00003
[5]  
CAMERON JR, 1967, NONLINEARITY THERMOL
[6]   DOSE-RATE DETERMINATION FOR I-125 SEEDS [J].
CHIUTSAO, ST ;
ANDERSON, LL ;
OBRIEN, K ;
SANNA, R .
MEDICAL PHYSICS, 1990, 17 (05) :815-825
[7]  
DEWERD LA, 1979, TOPICS APPLIED PHYSI
[8]   EFFECTS OF ANNEALING AND COOLING PROCESSES ON THERMOLUMINESCENCE OF LIF (TLD100) [J].
DHAR, A ;
DEWERD, LA ;
STOEBE, TG .
HEALTH PHYSICS, 1973, 25 (04) :427-433
[9]   SENSITIVITY ENHANCEMENT EFFECTS IN THE THERMO-LUMINESCENCE OF LIF TLD-100 AT RADIATION-DOSES BELOW 10 GY [J].
FOLKARD, M ;
ROPER, MJ ;
MICHAEL, BD .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06) :769-773
[10]   THE 3-DIMENSIONAL DOSE DISTRIBUTION OF I-125 SEEDS IN TISSUE [J].
HARTMANN, GH ;
SCHLEGEL, W ;
SCHARFENBERG, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (06) :693-699