SEMIQUANTITATIVE ANALYSIS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:33
作者
AMARASIRIWARDENA, CJ
GERCKEN, B
ARGENTINE, MD
BARNES, RM
机构
[1] Department of Chemistry, University of Massachusetts, GRC Towers, Amherst
关键词
Biological standard reference material; Inductively coupled plasma mass spectrometry; Semi-quantitative analysis; Trace element;
D O I
10.1039/ja9900500457
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Capabilities and applicability of a semi-quantitative analysis program were critically investigated with National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 1643b (Trace Elements in Water). Several instrument parameters and methods of analysis were examined in order to determine their effects on precision and accuracy. The best results were obtained through the division of the provided element response table into three groups and revision of the response table with V, Cs and Ir as internal references. Complex samples including biological and botanical SRMs were examined. Analytical results of 23 elements indicated that the elemental concentrations were within 30% of the certified values.