DETERMINATION OF MERCURY-VAPOR IN AIR USING ELECTRICALLY HEATED, GOLD-COATED WIRE-LOOPS, IN-TORCH VAPORIZATION SAMPLE INTRODUCTION AND INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY WITH PHOTODIODE-ARRAY DETECTION

被引:17
作者
KARANASSIOS, V [1 ]
BATEMAN, KP [1 ]
SPIERS, GA [1 ]
机构
[1] UNIV GUELPH,GUELPH N1G 2W1,ON,CANADA
关键词
D O I
10.1016/0584-8547(94)80076-6
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A method for the determination of Hg-0 vapor in ambient air using an electrically heated, Au-coated wire-loop, in-torch vaporization (ITV) sample introduction and inductively coupled plasma atomic emission spectrometry (ICP-AES) spectrometry with photodiode array detection is described. Tungsten wire-loops were coated with gold using sputtering methodology developed for scanning electron microscopy (SEM). Mecury vapor formed an amalgam with Au on the wire-loop, the wire-loop was inserted into a modified ICP torch about 10 cm below the plasma and electrical power was applied to the loop, thus thermally releasing mercury. Atomic emission was monitored using an ICP spectrometer equipped with a photodiode array detector. Calibration curves were constructed using a variation of the reduction-aeration method and were linear for about two-and-a-half orders of magnitude. Mercury vapor levels determined close to a polarography apparatus compared favorably with measurements taken at the same location using cold vapor atomic absorption spectrometry (CVAAS). The possibility of using a Au-coated wire-loop system as an integrating personal dosimeter is postulated.
引用
收藏
页码:867 / 874
页数:8
相关论文
共 18 条
[1]   DETERMINATION OF VOLATILE MERCURY SPECIES AT THE PICOGRAM LEVEL BY LOW-TEMPERATURE GAS-CHROMATOGRAPHY WITH COLD-VAPOR ATOMIC FLUORESCENCE DETECTION [J].
BLOOM, N ;
FITZGERALD, WF .
ANALYTICA CHIMICA ACTA, 1988, 208 (1-2) :151-161
[2]   DETERMINATION OF SMALL AMOUNTS OF MERCURY [J].
CHILOV, S .
TALANTA, 1975, 22 (03) :205-232
[3]  
DANIELS RS, 1991, ANAL CHIM ACTA, V248, P575
[4]   COMPARISON OF THE COLLECTION AND DESORPTION EFFICIENCY OF ACTIVATED-CHARCOAL, SILVER, AND GOLD FOR THE DETERMINATION OF VAPOR-PHASE ATMOSPHERIC MERCURY [J].
DUMAREY, R ;
DAMS, R ;
HOSTE, J .
ANALYTICAL CHEMISTRY, 1985, 57 (13) :2638-2643
[5]   A NEW CALIBRATION METHOD FOR THE DETERMINATION OF TRACE AMOUNTS OF MERCURY IN AIR AND BIOLOGICAL-MATERIALS [J].
FRIESE, KH ;
ROSCHIG, M ;
WUENSCHER, G ;
MATSCHINER, H .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (08) :860-866
[6]   DETERMINATION OF SUB-MICROGRAM QUANTITIES OF MERCURY BY ATOMIC ABSORPTION SPECTROPHOTOMETRY [J].
HATCH, WR ;
OTT, WL .
ANALYTICAL CHEMISTRY, 1968, 40 (14) :2085-&
[7]   DISTRIBUTION OF ATMOSPHERIC MERCURY SPECIES NEAR GROUND [J].
JOHNSON, DL ;
BRAMAN, RS .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1974, 8 (12) :1003-1009
[8]   ELECTRICALLY HEATED WIRE-LOOP, IN-TORCH VAPORIZATION SAMPLE INTRODUCTION SYSTEM FOR INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY WITH PHOTODIODE-ARRAY DETECTION [J].
KARANASSIOS, V ;
BATEMAN, KP ;
SPIERS, GA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (09) :847-865
[9]   NON-FLAME ATOMIC ADSORPTION ATTACHMENT FOR TRACE MERCURY DETERMINATION [J].
LIEU, VT ;
CANNON, A ;
HUDDLESTON, WE .
JOURNAL OF CHEMICAL EDUCATION, 1974, 51 (11) :752-753
[10]  
LUCA C, 1987, REV ANAL CHEM, V9, P1