PRECISION OF COLOR MEASUREMENT WITH GE SPECTROPHOTOMETER .2. PHOTOMETRIC ACCURACY

被引:5
作者
RHODES, EC
BILLMEYER, FW
机构
[1] Rensselaer Polytechnic Institute, Packard Electric Division, General Motors Corporation, Troy, NY
[2] Rensselaer Polytechnic Institute, Troy, NY
关键词
D O I
10.1364/AO.8.000769
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The supplementary light method of Reule was applied to calibrate a simple phototube photometer to an absolute accuracy of ±0.006% (95% confidence limits). The photometer was then used to determine the photometric accuracy of a General Electric recording spectrophotometer. The photometric scale of this instrument was shown to be accurate to within 0.1% over its entire range. © 1969 Optical Society of America.