EXPERIMENTAL EVALUATION OF 7 DIFFERENT SPINAL FRACTURE INTERNAL-FIXATION DEVICES USING NONFAILURE STABILITY TESTING - THE LOAD-SHARING AND UNSTABLE-MECHANISM CONCEPTS

被引:44
作者
GAINES, RW
CARSON, WL
SATTERLEE, CC
GROH, GI
机构
[1] Division of Orthopaedic Surgery, University of Missouri, Columbia, MO
[2] Department of Mechanical and Aerospace Engineering, University of Missouri Columbia, Columbia, MO
关键词
3-DIMENSIONAL BIOMECHANICAL EVALUATION; SPINE FIXATION; LOAD SHARING; STABILITY; UNSTABLE MECHANISM;
D O I
10.1097/00007632-199108000-00007
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Fracture site immobilization capabilities of seven internal spine fixation systems were experimentally evaluated: Harrington double distraction (plain, supplemented with Edwards sleeves, supplemented with sublaminar wires), Luque rectangle, Kaneda device, transfixed Kaneda device, and Steffee plates. Stability evaluation involved comparing the three-dimensional motion that occurred across an experimentally created L1 slice fracture, and the load resistance of the implants when subjected to axial, flexion, extension, lateral bending, and torsional loads. Each implant was tested on 15 different vertebral segments from 200-250-lb calves. All implants load-shared with the fractured vertebral column to varying degrees. All except the Steffee plate system showed an obvious biomechanical weakness in one or more load modes. The unstable 4R-4bar mechanism configuration of bilevel spinal implants was identified. Mechanism configurations allow free movement with little or no resistance to the applied load until load sharing by the spinal column stops the collapse and resists the applied load.
引用
收藏
页码:902 / 909
页数:8
相关论文
共 29 条
[1]  
Asher M., Carson W.L., Heinig C., A modular spinal rod linkage system to provide rotational stability, Spine, 13, pp. 272-277, (1988)
[2]  
Ashman R.B., Galpin R.D., Corin J.D., Johnson C.E., Biomechanical analysis o f pedicle screw instrumentation systems in a corpectomy model, Spine, 14, pp. 1398-1405, (1989)
[3]  
Ashman R.B., Birch J.G., Bone L.B., Mechanical testing o f spinal instrumentation, Clin Orthop, 227, pp. 113-125, (1988)
[4]  
Carson W.L., Duffield R.C., Arendt M., Ridgely B.J., Gaines R.W., Internal Forces and Moments in Transpedicular Spine Instrumentation: The Effect of Pedicle Screw Angle and transfixation—The 4R-4Bar Linkage Concept. Presented at The Joint Meeting of The Scoliosis Research Society, Amsterdam, (1989)
[5]  
Courtney S.P., Saha S., Lipka J.M., Albright J.A., A Biomechanical Analysis of the Steffee Plated Thoracolumbar Spine. Proceedings of the World Congress on Medical Physics and Biomedical Engineers, pp. 1-4, (1988)
[6]  
Ferguson R.L., Tencer A.F., Woodward P., Allen B.L., Biomechanical comparisons o f spinal fracture models and the stabilizing effects o f posterior instrumentations, Spine, 13, pp. 453-460, (1988)
[7]  
Fidler M.W., Posterior instrumentation of the spine: An experimental comparison of various possible techniques, Spine, 11, pp. 367-371, (1986)
[8]  
Gaines R.W., Carson W.L., Huibregtse D.M., Biomechanics O F Thoracic and Thoracolumbar Fractures: A Comprehensive Re-Evaluation, (1986)
[9]  
Gaines R.W., Carson W.L., Pak D.J., Huibregtse D.M., Humphreys W.G., Poster 1 7 3, presented at the orthopaedic research society 32nd, Comparison of Harrington Rods Alone, Harrington Rods with Edwards Sleeves, and Harrington Rods with Sublaminar Wires: Experimental Investigation in a Calf Spine Model, (1986)
[10]  
Gaines R.W., Breedlove R.F., Munson G., Stabilization of thoracic and thoracolumbar fracture: Dislocations with harrington rods and sublaminar wires, Clin Orthop, 189, pp. 195-203, (1984)