IS STEM LENGTH IMPORTANT IN UNCEMENTED ENDOPROSTHESES

被引:17
作者
TANNER, KE
YETTRAM, AL
LOEFFLER, M
GOODIER, WD
FREEMAN, MAR
BONFIELD, W
机构
[1] BRUNEL UNIV,DEPT MECH ENGN,UXBRIDGE UB8 3PH,MIDDX,ENGLAND
[2] ROYAL LONDON HOSP,DEPT ORTHOPAED SURG,LONDON E1 1BB,ENGLAND
关键词
STEM LENGTH; UNCEMENTED ENDOPROSTHESES; FINITE ELEMENT ANALYSIS;
D O I
10.1016/1350-4533(95)90854-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
In an uncemented total hip replacement where the femoral stem is thinner than the medullary canal it may be hypothesized that the distal portion of the stem has no mechanical function. In this study an uncemented total hip replacement has been modelled mathematically using finite element analysis and mechanical tests of a similar system have been carried out loading implants in cadaveric proximal femora. Two implants have been tested mechanically; one with a full length stem and the second with the stem shortened, but three implants have been modelled with an additional intermediate length stem analysed. Additionally the finite element analysis has been done with a high neck resection and a standard level resection. The finite element model showed that the full length sta produced lower interface contact stress levels under the proximal neck, particularly when the neck was resected, and this was borne out by the mechanical testing.
引用
收藏
页码:291 / 296
页数:6
相关论文
共 9 条
[1]
ALBREKTSSON B E J, 1987, Journal of Orthopaedic Surgical Techniques, V3, P99
[2]
WHY RESECT THE NECK [J].
FREEMAN, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (03) :346-349
[3]
GEESINK RGT, 1987, CLIN ORTHOP RELAT R, V225, P147
[4]
QUANTIFICATION OF BONE STRESSES DURING REMODELING [J].
HASSLER, CR ;
RYBICKI, EF ;
CUMMINGS, KD ;
CLARK, LC .
JOURNAL OF BIOMECHANICS, 1980, 13 (02) :185-190
[5]
LIGGINS AB, 1992, EXPT MECH TECHNOLOGY, P61
[6]
EFFECTS OF STEM DESIGN AND MATERIAL PROPERTIES ON STRESSES IN HIP ENDOPROSTHESES [J].
ROHLMANN, A ;
MOSSNER, U ;
BERGMANN, G ;
HEES, G ;
KOLBEL, R .
JOURNAL OF BIOMEDICAL ENGINEERING, 1987, 9 (01) :77-83
[7]
A SYSTEM FOR MODELING FORCES ON THE HIP-JOINT IN ONE-LEGGED STANCE [J].
TANNER, KE ;
REED, PE ;
BONFIELD, W ;
RASMUSSEN, GL ;
FREEMAN, MAR .
JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (03) :289-290
[8]
TRENDS OF MECHANICAL CONSEQUENCES AND MODELING OF A FIBROUS MEMBRANE AROUND FEMORAL HIP PROSTHESES [J].
WEINANS, H ;
HUISKES, R ;
GROOTENBOER, HJ .
JOURNAL OF BIOMECHANICS, 1990, 23 (10) :991-1000
[9]
EFFECT OF INTERFACE CONDITIONS ON THE BEHAVIOR OF A FREEMAN HIP ENDOPROSTHESIS [J].
YETTRAM, AL .
JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (06) :520-524