PERFORMANCE-CHARACTERISTICS OF THE KIN-COM(R) DYNAMOMETER

被引:50
作者
MAYHEW, TP
ROTHSTEIN, JM
FINUCANE, SDG
LAMB, RL
机构
[1] UNIV ILLINOIS,COLL ASSOCIATED HLTH PROFESSIONS,DEPT PHYS THERAPY,CHICAGO,IL 60612
[2] UNIV ILLINOIS HOSP,PHYS THERAPY SERV,CHICAGO,IL 60612
来源
PHYSICAL THERAPY | 1994年 / 74卷 / 11期
关键词
DEVICE TESTING; EQUIPMENT; TESTS AND MEASUREMENTS;
D O I
10.1093/ptj/74.11.1047
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and Purpose. The purpose of this study was to assess the performance characteristics of a Kin-Com(R) dynamometer (model #500-11) under controlled conditions. Methods. Comparisons were made between measurments of force, angle, and velocity obtained from the Kin-Com(R) and measurements acquired from an external recording system of known weights, angles, and user-set velocities. The strength of the linear relationships between measurements obtained with the different recording systems was analyzed using a coefficient of determination (r2). An intraclass correlation coefficient (ICC[2,1]) was used to examine the reliability of the force, angle, and velocity measurements obtained with each recording system on 2 different days. Results. In all conditions, the coefficient of determination for the force, angle, and velocity comparisons was above .99. The ICC for between-day comparisons for all force, angle, and velocity measurements was above .99. Conclusion and Discussion. Our results indicate that the static measurements of force and angle that are necessary for use in the gravity-correction procedure and isometric testing are accurate and replicable between days. The Kin-Com(R) dynamometer's control system regulating lever arm velocity is also accurate and replicable under a no-load condition. It was ascertained during the velocity testing that the use of any acceleration and deceleration mode other than ''high'' resulted in a loss of excursion of the lever arm.
引用
收藏
页码:1047 / 1054
页数:8
相关论文
共 16 条
[1]  
AITKENS S, 1987, PHYS THER, V67, P756
[2]  
FARRELL M, 1986, MED SCI SPORT EXER, V18, P44
[3]   IMPORTANCE OF CORRECTING ISOKINETIC PEAK TORQUE FOR THE EFFECT OF GRAVITY WHEN CALCULATING KNEE FLEXOR TO EXTENSOR MUSCLE RATIOS [J].
FILLYAW, M ;
BEVINS, T ;
FERNANDEZ, L .
PHYSICAL THERAPY, 1986, 66 (01) :23-29
[4]  
FRANCIS K, 1987, Journal of Orthopaedic and Sports Physical Therapy, V8, P480
[5]  
HAGEMAN P A, 1988, Journal of Orthopaedic and Sports Physical Therapy, V10, P59
[6]   EFFECT OF STABILIZATION ON MAXIMAL ISOKINETIC TORQUE OF THE QUADRICEPS FEMORIS MUSCLE DURING CONCENTRIC AND ECCENTRIC CONTRACTIONS [J].
HANTEN, WP ;
RAMBERG, CL .
PHYSICAL THERAPY, 1988, 68 (02) :219-222
[7]   EFFECT OF COOLING ON FORCE OSCILLATIONS DURING MAXIMAL VOLUNTARY ECCENTRIC EXERCISE [J].
HART, DL ;
MILLER, LC ;
STAUBER, WT .
EXPERIMENTAL NEUROLOGY, 1985, 90 (01) :73-80
[8]   EVALUATION OF ECCENTRIC EXERCISE IN TREATMENT OF PATELLAR TENDINITIS [J].
JENSEN, K ;
DIFABIO, RP .
PHYSICAL THERAPY, 1989, 69 (03) :211-216
[9]  
LORD J, 1987, PHYS THER, V67, P757
[10]  
Mayhew T, 1985, MEASUREMENT PHYSICAL, P57