Ion mobility spectrometry peak width data are fitted by a least-squares procedure to a semiempirical model having three adjustable parameters. Peaks are wider than contributions from initial pulse width and diffusion predict, and it is suggested that the additional width is due mainly to electric field inhomogeneity and Coulombic repulsion. The effects of operating conditions and instrument dimensions on resolving power are discussed. It is proposed that increased inhomogeneity of the electric field results in lower measured mobility values, as well as lower resolving power.
引用
收藏
页码:4195 / 4201
页数:7
相关论文
共 11 条
[1]  
ALBRITTON DL, 1967, MEASUREMENT LOW ENER
[2]  
ARONSON EA, 1987, SAND870072 SAND CORP
[3]  
BROKENSHIRE JL, 1991, OCT FACSS M AN
[4]   ION MOBILITY SPECTROMETRY [J].
HILL, HH ;
SIEMS, WF ;
STLOUIS, RH ;
MCMINN, DG .
ANALYTICAL CHEMISTRY, 1990, 62 (23) :A1201-A1209
[5]  
MCDANIEL EW, 1964, COLLISION PHENOMENA, pCH11
[6]   THEORY OF PLASMA CHROMATOGRAPHY GASEOUS ELECTROPHORESIS - REVIEW [J].
REVERCOMB, HE ;
MASON, EA .
ANALYTICAL CHEMISTRY, 1975, 47 (07) :970-983
[7]  
ROKUSHIKA S, 1975, ANAL CHEM, V57, P1902
[8]   TABLE OF REDUCED MOBILITY VALUES FROM AMBIENT PRESSURE ION MOBILITY SPECTROMETRY [J].
SHUMATE, C ;
STLOUIS, RH ;
HILL, HH .
JOURNAL OF CHROMATOGRAPHY, 1986, 373 (02) :141-173
[9]   PEAK SHAPE ANALYSIS AND PLATE THEORY FOR PLASMA CHROMATOGRAPHY [J].
SPANGLER, GE ;
COLLINS, CI .
ANALYTICAL CHEMISTRY, 1975, 47 (03) :403-407
[10]   SPACE-CHARGE EFFECTS IN ION MOBILITY SPECTROMETRY [J].
SPANGLER, GE .
ANALYTICAL CHEMISTRY, 1992, 64 (11) :1312-1312