VALIDATION OF QUANTITATIVE CHROMATOGRAPHIC ANALYSIS ON LABORATORY-PREPARED THIN-LAYERS

被引:7
作者
PETROVIC, M [1 ]
KASTELANMACAN, M [1 ]
机构
[1] UNIV ZAGREB, FAC CHEM ENGN & TECHNOL, ANALYT CHEM LAB, HR-10002 ZAGREB, CROATIA
关键词
D O I
10.1016/0021-9673(95)00018-I
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Validation of the determination of amino acids on laboratory-prepared plates was performed on a mixed natural zeolite and microcrystalline cellulose sorbent. Chromatograms were recorded in the visible region using a Datacolor DC 3890 reflectance spectrometer, Camag Turner Fluorimeter 111 slit-scanning densitometer and Leco 2001 image analyser. The precision, detection limit and limit of quantification for each method were determined. A lower total R.S.D. was obtained by measuring colour remission with the Datacolor system (2.7% for high concentration and 3% for low concentration), with an instrumental R.S.D. of 0.1-0.2%. Densitometry and image analysis resulted in a lower precision, especially at low concentration (total R.S.D.>10%), with a significant instrumental error.
引用
收藏
页码:173 / 178
页数:6
相关论文
共 4 条
[1]   AN IMPROVED METHOD OF DENSITOMETRIC THIN LAYER CHROMATOGRAPHY AS APPLIED TO DETERMINATION OF SAPOGENIN IN DIOSCOREA TUBERS [J].
BRAIN, KR ;
HARDMAN, R .
JOURNAL OF CHROMATOGRAPHY, 1968, 38 (03) :355-&
[2]  
MULLIN BR, 1985, TECHNIQUES APPLICATI
[3]   INSTRUMENTAL THIN-LAYER CHROMATOGRAPHY [J].
POOLE, CF ;
POOLE, SK .
ANALYTICAL CHEMISTRY, 1994, 66 (01) :A27-A37
[4]  
Poole CF., 1991, CHROMATOGRAPHY TODAY, VEdition