PRECISE SCANNING CALORIMETER FOR STUDYING THERMAL-PROPERTIES OF BIOLOGICAL MACROMOLECULES IN DILUTE-SOLUTION

被引:165
作者
PRIVALOV, G
KAVINA, V
FREIRE, E
PRIVALOV, PL
机构
[1] JOHNS HOPKINS UNIV,DEPT BIOL,BALTIMORE,MD 21218
[2] JOHNS HOPKINS UNIV,CTR BIOCALORIMETRY,BALTIMORE,MD 21218
关键词
D O I
10.1006/abio.1995.9957
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A precise scanning calorimeter for studying the heat capacity of liquids in a broad temperature range has been developed. By its design and capabilities this calorimeter is the first of a new generation for this type of instrument. This new scanning calorimeter operates differentially, is equipped with a pair of gold capillary cells and semiconductor sensors, and is able to scan up and down in temperature at user-selected rates. This instrument is completely operated by an integrated computer which also provides a full thermodynamic analysis of the results. Its construction does not involve the use of organic compounds, thus eliminating a source of baseline noise that has affected previous calorimeters. The operational temperature range of the instrument can be varied between 0 and 120 degrees C. The gold capillary cells (operational volume 0.8 ml) minimize temperature gradients in the heated/cooled liquid sample and permit easy washing and reloading without air bubbles. These features are crucial for the accuracy of difference heat capacity measurements and determination of the absolute value of the partial heat capacity of solute molecules. The measurements can be performed under an excess constant pressure (up to 3 atm) to prevent formation of gas bubbles and boiling of aqueous solutions above 100 degrees C. The noise level of the recorded heating/cooling power difference is below 50 x 10(-9) W (i.e., below 10 ncal/s) with a response half-time of 5 s. The reproducibility of the baseline without refilling the capillary cells is on the order of 0.5 x 10(-6) W. Reproducibility of the baseline upon refilling the cell is of the same order of magnitude. This provides an accuracy in difference heat capacity determination on the order of 10 mu cal/degrees Kml at a heating rate of 1 degrees K/min. (C) 1995 Academic Press, Inc.
引用
收藏
页码:79 / 85
页数:7
相关论文
共 21 条
[1]   DIFFERENTIAL CALORIMETRY OF THERMALLY INDUCED PROCESSES IN SOLUTION [J].
DANFORTH, R ;
KRAKAUER, H ;
STURTEVANT, JM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1967, 38 (04) :484-+
[2]  
Freire E, 1995, METHOD ENZYMOL, V259, P144
[3]   STATISTICAL MECHANICAL DECONVOLUTION OF THERMAL TRANSITIONS IN MACROMOLECULES .1. THEORY AND APPLICATION TO HOMOGENEOUS SYSTEMS [J].
FREIRE, E ;
BILTONEN, RL .
BIOPOLYMERS, 1978, 17 (02) :463-479
[4]   DIFFERENTIAL HEAT CAPACITY CALORIMETER FOR POLYMER TRANSITION STUDIES [J].
GILL, SJ ;
BECK, K .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1965, 36 (03) :274-&
[5]   THE HEAT-CAPACITY OF PROTEINS [J].
GOMEZ, J ;
HILSER, VJ ;
XIE, D ;
FREIRE, E .
PROTEINS-STRUCTURE FUNCTION AND GENETICS, 1995, 22 (04) :404-412
[6]   ENERGETICS OF THE ALPHA-LACTALBUMIN STATES - A CALORIMETRIC AND STATISTICAL THERMODYNAMIC STUDY [J].
GRIKO, YV ;
FREIRE, E ;
PRIVALOV, PL .
BIOCHEMISTRY, 1994, 33 (07) :1889-1899
[7]  
Privalov P L, 1979, Adv Protein Chem, V33, P167, DOI 10.1016/S0065-3233(08)60460-X
[8]  
Privalov P.L., 1964, J EXP THEOR PHYS+, V47, P2073
[9]   THERMODYNAMIC INVESTIGATIONS OF BIOLOGICAL MACROMOLECULES [J].
PRIVALOV, PL .
PURE AND APPLIED CHEMISTRY, 1976, 47 (04) :293-304
[10]   3 GENERATIONS OF SCANNING MICROCALORIMETERS FOR LIQUIDS [J].
PRIVALOV, PL ;
PLOTNIKOV, VV .
THERMOCHIMICA ACTA, 1989, 139 :257-277