INTER-LABORATORY COMPARISON OF BLOOD LEAD DETERMINATIONS

被引:11
作者
MAHER, CC
ROETTGERS, DM
CONLON, HJ
机构
[1] Industrial Toxicology Laboratory, West Allis Memorial Hospital, West Allis
来源
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL | 1979年 / 40卷 / 03期
关键词
D O I
10.1080/15298667991429525
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Results are presented for an interlaboratory proficiency study of blood lead determinations. Samples were pooled from individuals occupationally exposed to lead. Performance of individual laboratories is compared over a two year period. With increased emphasis in recent years on proficiency studies of this type, agreement between laboratories has been somewhat improved, but blood remains a difficult matrix in which to measure lead concentration. Copyright 1979, American Industrial Hygiene Association
引用
收藏
页码:230 / 237
页数:8
相关论文
共 16 条
  • [1] BERITIC T, 1977, J OCCUP ENVIRON MED, V19, P551
  • [2] BERLIN A, 1973, ENV HLTH ASPECTS LEA, P1033
  • [3] BROWNE RC, 1974, LANCET, V2, P1112
  • [4] LABORATORIES WHICH CONDUCT LEAD ANALYSES ON BIOLOGIC SPECIMENS
    DONOVAN, DT
    VOUGHT, VM
    RAKOW, AB
    [J]. ARCHIVES OF ENVIRONMENTAL HEALTH, 1971, 23 (02): : 111 - &
  • [5] ECKERT AC, 1976, NBS422 SPEC PUBL, P275
  • [6] GOYER RA, 1977, TOXICOLOGY TRACE ELE, P41
  • [7] HERNBERG S, 1970, LANCET, V1, P63
  • [8] KEENAN R G, 1963, Am Ind Hyg Assoc J, V24, P481, DOI 10.1080/00028896309343251
  • [9] INTERLABORATORY EVALUATION OF RELIABILITY OF BLOOD LEAD ANALYSES
    KEPPLER, JF
    MAXFIELD, ME
    MOSS, WD
    TIETJEN, G
    LINCH, AL
    [J]. AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1970, 31 (04): : 412 - &
  • [10] LAMOLA AA, 1975, CLIN CHEM, V21, P93