A FLOW-THROUGH MICROSAMPLING DEVICE APPLIED TO AN ION-EXCHANGE CHROMATOGRAPHY SYSTEM

被引:16
作者
NILSSON, J
SZECSI, P
SCHAFERNIELSEN, C
机构
[1] MALMO GEN HOSP,DEPT CLIN CHEM,S-21401 MALMO,SWEDEN
[2] GLOSTRUP CTY HOSP,DEPT CLIN CHEM,DK-2600 GLOSTRUP,DENMARK
来源
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS | 1993年 / 27卷 / 03期
关键词
PIEZOELECTRIC DROP GENERATOR; FLOW-THROUGH LIQUID SAMPLER; FUSED ROCKET ELECTROPHORESIS;
D O I
10.1016/0165-022X(93)90002-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A flow-through microsampling device that can be inserted in the column elution flow line of a chromatographic system has been developed. A sample drop (volume 2.2 nl) is ejected when a voltage pulse is applied to a piezo-electric transducer. The drops can be ejected with up to 70 Hz frequency, giving a sample flow range of 0-9 mul/min. Samples were collected on a moving agarose surface during elution of freshly prepared human serum. The precipitate pattern obtained after immunoelectrophoresis is similar to what would be obtained in a fused rocket electrophoresis of multiple samples from a fraction collector.
引用
收藏
页码:181 / 190
页数:10
相关论文
共 11 条
  • [1] Axelsen N. H., 1973, SCAND J IMMUNOL S1, V2, P1
  • [2] BENTIN H, 1986, J IMAGING TECHNOL, V12, P152
  • [3] BROCK JD, 1984, J IMAGING TECHNOL, V10, P127
  • [4] GOOBERMAN GL, 1968, ULTRASONICS THEORY A, P22
  • [5] HARA T, 1981, BUBBLE JET RECORDING
  • [6] INK-JET PRINTING
    HEINZL, J
    HERTZ, CH
    [J]. ADVANCES IN IMAGING AND ELECTRON PHYSICS, 1985, 65 : 91 - 171
  • [7] CLEAVAGE OF STRUCTURAL PROTEINS DURING ASSEMBLY OF HEAD OF BACTERIOPHAGE-T4
    LAEMMLI, UK
    [J]. NATURE, 1970, 227 (5259) : 680 - +
  • [8] LLOYD WJ, 1987, OUTPUT HARDCOPY DEVI, P331
  • [9] OFARELL PJ, 1973, SCAND J IMMUNOL S1, V2, P69
  • [10] Stemme N.G.E., 1973, U.S. Patent, Patent No. [US3747120 A, 3747120, US3747120A]