THE USE OF H-3 STANDARDS IN I-125 AUTORADIOGRAPHY

被引:58
作者
ARTYMYSHYN, R
SMITH, A
WOLFE, BB
机构
[1] GEORGETOWN UNIV,SCH MED,DEPT PHARMACOL,3900 RESERVOIR RD NW,WASHINGTON,DC 20007
[2] UNIV PENN,SCH MED,DEPT PHARMACOL,PHILADELPHIA,PA 19104
关键词
Quantitative autoradiography; Standards;
D O I
10.1016/0165-0270(90)90139-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
To avoid repeated production of 125I tissue mash standards and to decrease inter-experiment variability, a set of commercially available, high activity plastic 3H standards was calibrated for use with 125I autoradiographs. Standards with a range of 2.0-7390 μCi 3H/g plastic were used since they produced autoradiographs of suitable optical density (darkness) at exposure times of 4 to more than 72 h. Tissue mash standards at 12 concentrations of 125I were produced by homogenizing rat brains and adding 125I. These were frozen and sectioned at thicknesses ranging from 6 to 120 μm and mounted on microscope slides. Each set of tissue standards, along with slides bearing the plastic 3H standards, was apposed to film for various times. The calibration of the 3H standards was done separately for each time point and section thickness. Autoradiographs were analyzed using a computer assisted image analysis system. The standard curve generated for each set of 125I-labelled tissue standards was correlated with the 3H standards. It was found that the relationship between the 3H plastic standards and the 125I tissue mash standards was not simple and depended on exposure time, thickness of the section, as well as the concentration of radioactivity in a given standard. However, this accurate calibration of the 3H standards yields long-lived standards for 125I autoradiography useful over a wide range of radionuclide concentrations, tissue thicknesses and exposure times. © 1990.
引用
收藏
页码:185 / 192
页数:8
相关论文
共 8 条
[1]   AN INEXPENSIVE MICROCOMPUTER DIGITAL IMAGING-SYSTEM FOR DENSITOMETRY - QUANTITATIVE AUTORADIOGRAPHY OF INSULIN-RECEPTORS WITH I-125 AND LKB ULTROFILM [J].
BASKIN, DG ;
DAVIDSON, D ;
CORP, ES ;
LEWELLEN, T ;
GRAHAM, M .
JOURNAL OF NEUROSCIENCE METHODS, 1986, 16 (02) :119-129
[2]   COMPARISON BETWEEN BRAIN PASTE AND POLYMER [I-125]STANDARDS FOR QUANTITATIVE RECEPTOR AUTORADIOGRAPHY [J].
DAVENPORT, AP ;
HALL, MD .
JOURNAL OF NEUROSCIENCE METHODS, 1988, 25 (01) :75-82
[3]  
FEINGOLD E, 1986, EXPT BIOL MED, V11, P175
[4]   QUANTITATIVE FILM AUTORADIOGRAPHY FOR TRITIUM - METHODOLOGICAL CONSIDERATIONS [J].
GEARY, WA ;
TOGA, AW ;
WOOTEN, GF .
BRAIN RESEARCH, 1985, 337 (01) :99-108
[5]  
LOWRY OH, 1951, J BIOL CHEM, V193, P265
[6]   THE USE OF C-14-LABELED TISSUE PASTE STANDARDS FOR THE CALIBRATION OF I-125 LABELED LIGANDS IN QUANTITATIVE AUTORADIOGRAPHY [J].
MILLER, JA ;
ZAHNISER, NR .
NEUROSCIENCE LETTERS, 1987, 81 (03) :345-350
[7]   QUANTITATIVE AUTORADIOGRAPHY OF BETA-1-ADRENERGIC AND BETA-2-ADRENERGIC RECEPTORS IN RAT-BRAIN [J].
RAINBOW, TC ;
PARSONS, B ;
WOLFE, BB .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1984, 81 (05) :1585-1589
[8]   QUANTITATIVE RECEPTOR AUTORADIOGRAPHY USING [H-3]LABELED ULTRAFILM - APPLICATION TO MULTIPLE BENZODIAZEPINE RECEPTORS [J].
UNNERSTALL, JR ;
NIEHOFF, DL ;
KUHAR, MJ ;
PALACIOS, JM .
JOURNAL OF NEUROSCIENCE METHODS, 1982, 6 (1-2) :59-73