THE EFFECT OF EXERCISE ON THE ATRIAL ELECTROGRAM VOLTAGE IN YOUNG-PATIENTS

被引:27
作者
ROSS, BA
ZEIGLER, V
ZINNER, A
WOODALL, P
GILLETTE, PC
机构
[1] MED UNIV S CAROLINA,DEPT PEDIAT,CHARLESTON,SC 29425
[2] INTERMED CORP,ANGLETON,TX
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 12期
关键词
ATRIAL ELECTROGRAM; EXERCISE; TREADMILL TESTING; CHILDREN;
D O I
10.1111/j.1540-8159.1991.tb06479.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Atrial electrogram sensing is an important function in active individuals with permanently implanted bipolar dual chamber pacing systems. We undertook to determine the effect of vigorous exercise on the atrial electrogram size in 11 children and young adults (average age 12 years). Using a telemetry signal through a handheld programming wand, nine tracings were completely and clearly recorded for analysis. Six patients had tined/passive fixation atrial leads and three patients had screw-in/active fixation lead systems. All leads were bipolar. The atrial electrogram size for each patient was measured at rest and at each minute of exercise. The atrial electrogram size decreased with exercise from a mean of 5.08 mV to 3.44 mV (range 0.9-4.25 mV) (P = 0.002). The 1.64 mV mean decrease represented a 33.8% reduction (range 19%-56%) (P < 0.001). There was no difference in the change in atrial electrogram size between the two lead types. Treadmill exercise testing with telemetric data of atrial electrograms showed a decrease in atrial electrogram size produced by exercise and may be helpful in determining appropriate atrial sensitivity settings in selected individuals. Because of the documented decrease in atrial electrogram size produced by exercise, we recommend obtaining maximal atrial electrograms at the time of implant and use of pacing systems that allow maximal flexibility in atrial sensing especially in athletically active individuals.
引用
收藏
页码:2092 / 2097
页数:6
相关论文
共 11 条
[1]  
ALONSOPULPON L, 1985, CARDIAC PACING ELECT, P1125
[2]   DECREASE IN CANINE ENDOCARDIAL AND EPICARDIAL ELECTROGRAM VOLTAGES WITH EXERCISE - IMPLICATIONS FOR PACEMAKER SENSING [J].
BRICKER, JT ;
WARD, KA ;
ZINNER, A ;
GILLETTE, PC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (04) :460-464
[3]   THE USE OF EXERCISE TESTING IN CHILDREN TO EVALUATE ABNORMALITIES OF PACEMAKER FUNCTION NOT APPARENT AT REST [J].
BRICKER, JT ;
GARSON, A ;
TRAWEEK, MS ;
SMITH, RT ;
WARD, KA ;
VARGO, TA ;
GILLETTE, PC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (05) :656-660
[4]  
EINTHOVEN W, 1908, ARCH GES PHYSL MENSC, V122, P571
[5]   ATRIAL SENSING PERFORMANCE OF AV UNIVERSAL PACEMAKERS DURING EXERCISE [J].
FROHLIG, G ;
BLANK, W ;
SCHWERDT, H ;
SEN, S ;
BETTE, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (01) :47-60
[6]   ATRIAL SIGNAL VARIATIONS AND PACEMAKER MALSENSING DURING EXERCISE - A STUDY IN THE TIME AND FREQUENCY-DOMAIN [J].
FROHLIG, G ;
SCHWERDT, H ;
SCHIEFFER, H ;
BETTE, L .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (04) :806-813
[7]   CARDIAC PACING AND PACEMAKERS .3. SENSING CARDIAC ELECTROGRAM [J].
FURMAN, S ;
HURZELER, P ;
DECAPRIO, V .
AMERICAN HEART JOURNAL, 1977, 93 (06) :794-801
[8]   EFFECT OF RATE AUGMENTATION AND ISOPROTERENOL ON THE AMPLITUDE OF ATRIAL AND VENTRICULAR ELECTROGRAMS [J].
ROSENHECK, S ;
SCHMALTZ, S ;
KADISH, AH ;
MORADY, F .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (01) :101-102
[9]   VALSALVA-INDUCED VARIATIONS IN THE INTRACARDIAC SIGNAL [J].
ROSENQVIST, M ;
LAGERGREN, H ;
STRANDBERG, H ;
EDHAG, O .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (06) :856-861
[10]   STABILITY OF PACING THRESHOLD, IMPEDANCE, AND R-WAVE AMPLITUDE AT REST AND DURING EXERCISE [J].
SCHUCHERT, A ;
KUCK, KH ;
BLEIFELD, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12) :1602-1608