ELIMINATION OF COPPER INTERFERENCE BY CONTINUOUS-FLOW MATRIX-ISOLATION IN THE DETERMINATION OF SELENIUM BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY

被引:29
作者
OFFLEY, SG
SEARE, NJ
TYSON, JF
KIBBLE, HAB
机构
[1] UNIV MASSACHUSETTS,DEPT CHEM,AMHERST,MA 01003
[2] LOUGHBOROUGH UNIV TECHNOL,DEPT CHEM,LOUGHBOROUGH LE11 3TU,LEICS,ENGLAND
[3] PHILIPS SCI,DEPT DEV,DIV ANALYT,CAMBRIDGE CB1 2PX,ENGLAND
关键词
SELENIUM DETERMINATION; FLOW INJECTION HYDRIDE GENERATION ATOMIC ABSORPTION SPECTROMETRY; COPPER INTERFERENCE REMOVAL; CONTINUOUS FLOW MATRIX ISOLATION;
D O I
10.1039/ja9910600133
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow system incorporating a microcolumn of strongly acidic cation-exchange resin (Dowex 50W) to achieve continuous flow matrix isolation was used to eliminate copper interference in the determination of selenium by flow injection hydride generation atomic absorption spectrometry. The microcolumn manifold used for the selective retention of the copper interferent was interfaced with the hydride generation manifold through a flow injection sample injection valve. The two manifolds were made independent of each other in order to achieve optimum performance characteristics for both the matrix isolation and hydride generation. Following removal of the copper, a 400-mu-l sample was injected into a water carrier stream. This was merged with hydrochloric acid and subsequently with sodium tetrahydroborate soution. After introduction of argon, the selenium hydride was separated by a glass U-tube separator and atomized by a tube-in-flame atomizer. The intermittent regeneration of the column with 1 mol dm-3 HCl enabled repeated matrix isolation without any loss in resin efficiency or the need for column repacking and gives the option for total automation. The procedure was validated through accurate analyses of two copper metal reference materials, National Institute of Standards and Technology Standard Reference Material 454 Unalloyed Copper Xl and Bundesanstalt fur Materialforschung und -prufung, Germany, Certified Reference Material 361 Copper, containing 479 and 36-mu-g g-1 of Se(IV), respectively. The system was found to have a characteristic concentration of 1.0 ng ml-1, limit of detection of 2.1 ng ml-1, relative standard deviation of 1.5% (10 ng ml-1 Se(IV), n = 12) and sample throughput of 51 h-1.
引用
收藏
页码:133 / 138
页数:6
相关论文
共 37 条
[1]  
ASTROM O, 1982, ANAL CHEM, V54, P190
[2]  
BAX D, 1988, SPECTROCHIM ACTA B, V3, P1349
[3]  
BEDARD M, 1976, CAN J SPECTROSC, V21, P64
[4]   ELIMINATION OF INTERFERENCES IN DETERMINATION OF ARSENIC AND ANTIMONY BY HYDRIDE GENERATION USING MOLECULAR EMISSION CAVITY ANALYSIS (MECA) [J].
BELCHER, R ;
BOGDANSKI, SL ;
HENDEN, E ;
TOWNSHEND, A .
ANALYST, 1975, 100 (1192) :522-523
[5]   INTERFERENCE REDUCTION BY L-CYSTINE IN THE DETERMINATION OF ARSENIC BY HYDRIDE GENERATION [J].
BOAMPONG, C ;
BRINDLE, ID ;
LE, XC ;
PIDWERBESKY, L ;
PONZONI, CMC .
ANALYTICAL CHEMISTRY, 1988, 60 (11) :1185-1188
[6]   THIOUREA AS A COMPLEXING AGENT FOR REDUCTION OF COPPER INTERFERENCE IN THE DETERMINATION OF SELENIUM BY HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY [J].
BYE, R ;
ENGVIK, L ;
LUND, W .
ANALYTICAL CHEMISTRY, 1983, 55 (14) :2457-2458
[7]   DESIGN AND OPERATION OF AN AUTOSAMPLER CONTROLLED FLOW-INJECTION PRECONCENTRATION SYSTEM FOR LEAD DETERMINATION BY FLAME ATOMIC-ABSORPTION SPECTROMETRY [J].
BYSOUTH, SR ;
TYSON, JF ;
STOCKWELL, PB .
JOURNAL OF AUTOMATIC CHEMISTRY, 1989, 11 (01) :36-39
[9]   COMBINATION OF FLOW-INJECTION TECHNIQUES WITH ATOMIC SPECTROMETRY IN AGRICULTURAL AND ENVIRONMENTAL-ANALYSIS [J].
FANG, ZL ;
XU, SK ;
WANG, X ;
ZHANG, SC .
ANALYTICA CHIMICA ACTA, 1986, 179 :325-340
[10]   SOME ATOMIC-ABSORPTION HYDRIDE GENERATION INTER-ELEMENT INTERFERENCE REDUCTION STUDIES UTILIZING ION-EXCHANGE RESINS [J].