ANALYTICAL TECHNIQUES FOR CELL FRACTIONS .14. USE OF DRAINAGE SYPHONS IN A FAST-ANALYZER CUVET-ROTOR

被引:28
作者
ANDERSON, NG
机构
关键词
D O I
10.1016/0003-2697(69)90104-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A cuvet-rotor, designated G-IIB, has been developed that contains small-bore syphons attached to each of the fifteen cuvets. Drainage of reactants from the transfer disc was found to be approximately 90% complete in 1 second and 99% complete after 2.5 seconds during acceleration to 1000 rpm. Mixing of protein-biuret reagent solution and 30% sucrose occurred in 1 second when a vacuum was used to pull air back through the syphons and cuvets. Drainage of the cuvets through the syphons was almost quantitative (>99%) when air pressure and one cycle from 400 rpm to 1000 rpm and back to 400 rpm was used. The interval between successive sets of analyses may therefore be 1 minute or less. A method of heat cycling to make the Telflon cuvet-spacer conform to the glass windows is presented. © 1969.
引用
收藏
页码:59 / &
相关论文
共 1 条
[1]   ANALYTICAL TECHNIQUES FOR CELL FRACTIONS .12. A MULTIPLE-CUVET ROTOR FOR A NEW MICROANALYTICAL SYSTEM [J].
ANDERSON, NG .
ANALYTICAL BIOCHEMISTRY, 1969, 28 (1-3) :545-&