AN IMPROVED TRANSIENT HOT-WIRE METHOD FOR STUDYING THERMAL TRANSPORT IN CONDENSED MATTER

被引:10
作者
TAKAHASHI, H [1 ]
HIKI, Y [1 ]
KOGURE, Y [1 ]
机构
[1] NISHI TOKYO UNIV,FAC SCI & ENGN,UENOHARA,YAMANASHI 40901,JAPAN
关键词
D O I
10.1063/1.1144636
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The transient hot-wire method for simultaneously measuring the thermal conductivity and thermal diffusivity of materials has been modified to improve the accuracy of the measurements. In this new experimental method, called the multicurrent method, several constant current values are used for heating the hot wire in order to determine accurately the thermal transport quantities. The details of the multicurrent method are precisely described, and examples of experiments using the method with pure water, electrolytic solution, and ionic conducting glasses are given. It is shown that this method can be used both for liquid and solid materials; a relatively small amount of the specimen material is sufficient for each experiment; the material can be insulating as well as highly ionic-conductive; the accuracies of the determined thermal conductivity and the thermal diffusivity values are, respectively, around +/-1% and +/-5%.
引用
收藏
页码:2901 / 2907
页数:7
相关论文
共 11 条
[1]   THE AXIAL-FLOW ERROR IN THE THERMAL-CONDUCTIVITY PROBE [J].
BLACKWELL, JH .
CANADIAN JOURNAL OF PHYSICS, 1956, 34 (04) :412-417
[2]  
CARLSAW HS, 1959, CONDUCTION HEAT SOLI
[3]   THEORY OF TRANSIENT HOT-WIRE METHOD FOR MEASURING THERMAL-CONDUCTIVITY [J].
HEALY, JJ ;
DEGROOT, JJ ;
KESTIN, J .
PHYSICA B & C, 1976, 82 (02) :392-408
[4]   AN EXPERIMENTAL-METHOD FOR STUDYING THERMAL TRANSPORT IN CONDENSED MATTER [J].
HIKI, Y ;
TAKAHASHI, H ;
KOGURE, Y .
PROCEEDINGS OF THE JAPAN ACADEMY SERIES B-PHYSICAL AND BIOLOGICAL SCIENCES, 1993, 69 (03) :51-54
[5]  
HIKI Y, 1993, PHONON SCATTERING CO, V7, P295
[6]  
HIKI Y, IN PRESS SOLID STATE
[7]  
KOGURE Y, 1973, JPN J APPL PHYS, V12, P814, DOI 10.1143/JJAP.12.814
[8]   LOW-TEMPERATURE THERMAL-DIFFUSIVITY MEASUREMENT BY LASER-FLASH METHOD [J].
KOGURE, Y ;
MUGISHIMA, T ;
HIKI, Y .
JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 1986, 55 (10) :3469-3478
[9]   SIMULTANEOUS MEASUREMENT OF THE THERMAL-CONDUCTIVITY AND THE THERMAL-DIFFUSIVITY OF LIQUIDS BY THE TRANSIENT HOT-WIRE METHOD [J].
NAGASAKA, Y ;
NAGASHIMA, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (02) :229-232
[10]   HEAT-CAPACITY AND THERMAL-CONDUCTIVITY FROM PULSED WIRE PROBE MEASUREMENTS UNDER PRESSURE [J].
SANDBERG, O ;
ANDERSSON, P ;
BACKSTROM, G .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1977, 10 (05) :474-477