STRATEGIES FOR OPTIMIZED APPLICATION OF ANNULAR-PHASED-ARRAY SYSTEMS IN CLINICAL HYPERTHERMIA

被引:45
作者
WUST, P
NADOBNY, J
FELIX, R
DEUFLHARD, P
LOUIS, A
JOHN, W
机构
[1] KONRAD ZUSE ZENTRUM,W-1000 BERLIN 31,GERMANY
[2] TECH UNIV BERLIN,FACHBEREICH MATH,W-1000 BERLIN 10,GERMANY
[3] TECH UNIV BERLIN,FACHBEREICH ELEKTROTECH,W-1000 BERLIN 10,GERMANY
关键词
HYPERTHERMIA TREATMENT PLANNING; OPTIMIZATION; ANNULAR-PHASED-ARRAY SYSTEM;
D O I
10.3109/02656739109004986
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A theoretical framework is presented for optimized heating of deep-seated tumours by phase and amplitude steering. The optimization problem for a specific tumour and perfusion case results in a functional dependency between power-level and maximum obtainable therapeutic efficiency. Different optimization criteria and strategies are outlined, which cause an increase of power or thermal dose in the tumour. Three tumour models (central pelvic tumour, eccentric abdominal tumour with or without necrosis) are analysed in detail. The simulation studies predict that appreciable parts of these tumours (50-100%) can be heated efficiently (42.5-43-degrees-C) within the range of available and clinically tolerated power levels (1-5 kW/m), if tumour perfusion is less than 20-25 ml/100 g min. Some improvements are obtained by increasing the number of independent channels (from four to eight) and by the application of time-dependent (complementary) power-deposition patterns.
引用
收藏
页码:157 / 173
页数:17
相关论文
共 48 条
[2]  
BANK RE, 1985, PLTMG USERS GUIDE
[3]  
BOAG A, 1988, IEEE T MICROW THEORY, P307
[4]   FAST-FOURIER-TRANSFORM METHOD FOR CALCULATION OF SAR DISTRIBUTIONS IN FINELY DISCRETIZED INHOMOGENEOUS MODELS OF BIOLOGICAL BODIES [J].
BORUP, DT ;
GANDHI, OP .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1984, 32 (04) :355-360
[6]  
DEWHIRST MW, 1984, CANCER RES, V44, P4772
[7]   REGIONAL HYPERTHERMIA WITH AN ANNULAR PHASED-ARRAY IN THE EXPERIMENTAL TREATMENT OF CANCER - REPORT OF WORK IN PROGRESS WITH A TECHNICAL EMPHASIS [J].
GIBBS, FA ;
SAPOZINK, MD ;
GATES, KS ;
STEWART, JR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (01) :115-119
[8]  
Herman G T, 1976, Comput Biol Med, V6, P273, DOI 10.1016/0010-4825(76)90066-4
[9]  
HORNSLETH S, 1989, HYPERTHERMIC ONCOLOG, V1, P684
[10]   REGIONAL HYPERTHERMIA FOR EXTENSIVE PELVIC TUMORS USING AN ANNULAR PHASED-ARRAY APPLICATOR - A FEASIBILITY STUDY [J].
HOWARD, GCW ;
SATHIASEELAN, V ;
KING, GA ;
DIXON, AK ;
ANDERSON, A ;
BLEEHEN, NM .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (708) :1195-1201