COMPREHENSIVE ANALYSIS OF DYNAMIC ELASTIC RESPONSE FEET - SEATTLE ANKLE LITE FOOT VERSUS SACH FOOT

被引:99
作者
LEHMANN, JF
PRICE, R
BOSWELLBESSETTE, S
DRALLE, A
QUESTAD, K
机构
[1] University of Washington, Department of Rehabilitation Medicine, Seattle, WA
来源
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION | 1993年 / 74卷 / 08期
关键词
BELOW-KNEE AMPUTATION; ENERGY EXPENDITURE; GAIT ANALYSIS; PROSTHETICS;
D O I
10.1016/0003-9993(93)90013-Z
中图分类号
R49 [康复医学];
学科分类号
100215 [康复医学与理疗学];
摘要
This study evaluated biomechanical and metabolic performance differences between two prosthetic foot designs in light of their mechanical properties. Ten unilateral below-knee amputee subjects, at least 1 year after amputation, capable of walking and running, were studied. Differences in heel and forefoot compliance explained differences in gait events and alignment. Increased efficiency of pushoff in the Seattle Ankle/Lite Foot exists as evidenced by the decrease loading on the opposite limb during double support and a less shortened step length on the sound side compared to the SACH foot. The natural frequency of oscillation for the prosthetic feet was determined to be too high to provide energy storage and release synchronized with kinematic requirements because neither metabolic cost savings nor differences in metabolic efficiency were found. Comfortable walking speed and the nadir of metabolic rate and efficiency are not different. Via accelerometer measurement, it was found that the more compliant and lossy SACH foot heel was less likely to transmit high frequency vibration. © 1993.
引用
收藏
页码:853 / 861
页数:9
相关论文
共 19 条
[1]
BARD G, 1959, Arch Phys Med Rehabil, V40, P415
[2]
BOCCARDI S, 1981, J BIOMECH, V14, P33
[3]
Burgess E M, 1985, J Rehabil Res Dev, V22, P75, DOI 10.1682/JRRD.1985.07.0075
[4]
BURGESS EM, 1983, ORTHOTICS PROSTHET, V37, P25
[5]
CONTOYANNIS B, 1987, B AUSTR VET AFFAIRS
[6]
FISHER SV, 1978, ARCH PHYS MED REHAB, V59, P124
[7]
BIOMECHANICAL ANALYSIS OF THE INFLUENCE OF PROSTHETIC FEET ON BELOW-KNEE AMPUTEE WALKING [J].
GITTER, A ;
CZERNIECKI, JM ;
DEGROOT, DM .
AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 1991, 70 (03) :142-148
[8]
GOH JCH, 1984, PROSTHET ORTHOT INT, V8, P147
[9]
JAMES KB, 1986, AM J PHYS MED REHAB, V65, P301
[10]
The use of spectral analysis to assess the performance of shock absorbing footwear [J].
Johnson, G.R. .
Engineering in Medicine, 1986, 15 (03) :117-122