DETERMINATION OF AMMONIUM AND L-GLUTAMINE IN HYBRIDOMA CELL-CULTURES BY SEQUENTIAL FLOW-INJECTION ANALYSIS

被引:23
作者
CAMPMAJO, C
CAIRO, JJ
SANFELIU, A
MARTINEZ, E
ALEGRET, S
GODIA, F
机构
[1] UNIV AUTONOMA BARCELONA,UNITAT ENGN QUIM,E-08193 BARCELONA,SPAIN
[2] UNIV AUTONOMA BARCELONA,UNIDAD QUIM ANALIT,E-08193 BARCELONA,SPAIN
关键词
FLOW INJECTION ANALYSIS; AMMONIUM; L-GLUTAMINE; SEQUENTIAL DETERMINATION; HYBRIDOMA CULTURE;
D O I
10.1007/BF00749614
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A flow injection analytical system based on a gas diffusion membrane module for ammonia and an ammonium flow-through potentiometric detector has been set up for measurement of L-glutamine and ammonium ions in hybridoma cell cultures. The main feature of the system is,that the same basic analytical concept and equipment is used in both measurements, the only difference being for the determination of L-glutamine, in which the sample flows through an immobilized glutaminase cartridge. The conditions to enable the performance of both analysis consecutively, avoiding potential interferences by unwanted deamination of other compounds in the samples, have been determined. Finally, the proposed system has been compared with reference analytical methods for batch hybridoma cell culture experiments.