HOW SAFE ARE CERAMIC HEADS AS HIP ENDOPROSTHESES - A SERIES OF 3 HEAD FRACTURES WITHIN 3 MONTHS

被引:13
作者
BURCKHARDT, A [1 ]
BERBERAT, C [1 ]
机构
[1] CANTONAL HOSP, ORTHOPED CLIN, OLTEN, SWITZERLAND
关键词
D O I
10.1007/BF00451877
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A report is given of a series of three ceramic head fractures involving articulation with polyethylene cups. No sure cause for the marked increase in occurrence could be found, as the heads were not from the same manufacturing lot, nor was there evidence of a material defect or an error in the implantation technique, Head fractures in the combination ceramic-polyethylene are reported very rarely because of the very low statistical probability. For this reason the present authors are continuing to use ceramic heads, but wish, in the interests of ''informed consent,'' to draw attention to the risk of head fractures.
引用
收藏
页码:215 / 219
页数:5
相关论文
共 16 条
[1]  
BOUTIN P, 1972, REV CHIR ORTHOP, V58, P229
[2]  
Dorre E., 1983, CERAMICS SURG, P61
[3]  
GIERSE H, 1989, ORTHOP PRAX, V6, P387
[4]   5 YEARS EXPERIENCE WITH CERAMIC-METAL-COMPOSITE HIP ENDOPROSTHESES .1. CLINICAL-EVALUATION [J].
GRISS, P ;
HEIMKE, G .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1981, 98 (03) :157-164
[5]   5 YEARS EXPERIENCE WITH CERAMIC-METAL-COMPOSITE HIP ENDOPROSTHESES .2. MECHANICAL EVALUATIONS AND IMPROVEMENTS [J].
HEIMKE, G ;
GRISS, P .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1981, 98 (03) :165-171
[6]   IMPLANT FRACTURES WITH CERAMIC HIP ENDOPROSTHESES [J].
HEISEL, J ;
SCHMITT, E .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1987, 125 (05) :480-490
[7]   CERAMIC HEAD BREAKAGE - A RARE COMPLICATION IN HIP-JOINT ARTHROPLASTY [J].
KERN, S ;
SCHREIBER, A ;
HILFIKER, B .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1990, 128 (05) :543-548
[8]  
MITTELMEIER H, 1984, ORTHOP PRAX, V8, P665
[9]  
PASCHEN U, 1986, THESIS U FRANKFURT A
[10]  
PEIRO A, 1991, J ARTHROPLASTY, V6, P372