NEW METHOD FOR CALIBRATING A GE DETECTOR BY USING ONLY ZERO TO 4 EFFICIENCY POINTS

被引:22
作者
GUNNINK, R
机构
[1] Lawrence Livermore National Laboratory, Livermore
关键词
D O I
10.1016/0168-9002(90)90807-I
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have developed a method whereby the efficiency of a coaxial germanium detector can be determined with an accuracy of a few percent for the energy region from 0.05 to 4.0 MeV by using only the specifications provided by the manufacturer. The accuracy can be further improved to within 1-2% by using one to four (or more) measured efficiency points. Our method also allows the detector efficiency-versus-energy curve to be applied to any source-to-detector distance.