A NEW REENTRANT IONIZATION-CHAMBER FOR THE CALIBRATION OF IR-192 HIGH-DOSE RATE SOURCES

被引:36
作者
GOETSCH, SJ
ATTIX, FH
DEWERD, LA
THOMADSEN, BR
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
[2] UNIV WISCONSIN,DEPT HUMAN ONCOL,MADISON,WI 53706
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1992年 / 24卷 / 01期
关键词
IR-192; HIGH-DOSE RATE AFTERLOADERS; BRACHYTHERAPY; REENTRANT IONIZATION CHAMBERS;
D O I
10.1016/0360-3016(92)91037-N
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A re-entrant (well-type) ionization chamber has been designed and fabricated at the University of Wisconsin for use with iridium-192 high dose-rate (HDR) remote after-loading brachytherapy devices. The chamber was designed to provide an ionization current of about 10(-8) ampere with a nominal 10 curie iridium-192 source. A narrow opening is provided into the sensitive volume of the chamber to insert a Nucletron MicroSelectron catheter, or catheters with similar diameters from other HDR manufacturers. The chamber exhibits a flat response (+/- 0.1%) for any source position within +/-4 mm of the chamber center. A .300 volt chamber bias yields a 99.96% ion collection efficiency. The chamber is capable of being calibrated directly with an iridium-192 source which has in turn been calibrated with thimble-type ion chambers. Reproducibility for readings in the current mode for 10 consecutive insertions of the MicroSelectron iridium-192 HDR source is within 0.02% or less. Two thimble chambers calibrated by the U.S. National Institute of Standards and Technology provide calibration traceability of iridium-192 HDR sources and re-entrant chambers to a primary national standards laboratory. Results of activity measurements of 6 commercial iridium-192 HDR sources are reported.
引用
收藏
页码:167 / 170
页数:4
相关论文
共 5 条
[1]  
EZZELL G, 1989, SELECTRON USERS NEWS, P3
[2]  
GOETSCH SJ, 1991, MED PHYS, V18, P142
[3]   STANDARDIZATION OF IR-192 GAMMA-RAY SOURCES IN TERMS OF EXPOSURE [J].
LOFTUS, TP .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1980, 85 (01) :19-25
[4]  
1985, NCRP58 NAT COUNC RAD
[5]  
1976, ICRU24 INT COMM RAD