LONGITUDINAL IMAGE DEBLURRING IN SPIRAL CT

被引:28
作者
SCHLUETER, FJ
WANG, G
HSIEH, PS
BRINK, JA
BALFE, DM
VANNIER, MW
机构
[1] Mallinckrodt Institute of Radiology, Washington University, School of Medicine, St Louis, MO 63110
关键词
COMPUTED TOMOGRAPHY (CT); HELICAL TECHNOLOGY; IMAGE PROCESSING; PHYSICS;
D O I
10.1148/radiology.193.2.7972755
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To assess the feasibility of digital deconvolution techniques to improve longitudinal resolution of spiral computed tomography (CT) multiplanar reformations and evaluate how technical factors in deconvolution affect longitudinal resolution, noise, and edge ringing. MATERIALS AND METHODS: Longitudinal line spread function (LSF) of the system was estimated from longitudinal reformations of trans-axial spiral CT images of a step test phantom. By using the estimated LSF, longitudinal reformations of the phantom and three clinical spiral CT studies were deconvolved by the methods of Wiener filtering and constrained iterative deconvolution. Edge ringing and image noise were quantified for Wiener filtering and constrained iterative deconvolution. RESULTS: Longitudinal reformations were substantially deblurred and resolution improved after deconvolution. Anatomic boundaries in clinical images were more clearly delineated after restoration. The methods of Wiener deconvolution and constrained iterative deconvolution improved the sharpness of the phantom step boundary at the expense of increased edge ringing and image noise. CONCLUSION: In longitudinal spiral CT reformations, blurring along the longitudinal axis can be reduced by Wiener filtering or constrained iterative deconvolution.
引用
收藏
页码:413 / 418
页数:6
相关论文
共 25 条
  • [1] QUANTITATIVE-ANALYSIS OF ELECTROPHORETOGRAMS - A MATHEMATICAL APPROACH TO SUPER-RESOLUTION
    AGARD, DA
    STEINBERG, RA
    STROUD, RM
    [J]. ANALYTICAL BIOCHEMISTRY, 1981, 111 (02) : 257 - 268
  • [2] AGARD DA, 1989, METHOD CELL BIOL, V30, P353
  • [3] AGARD DA, 1984, ANNU REV BIOPHYS BIO, V13, P191
  • [4] ITERATIVE METHODS FOR IMAGE DEBLURRING
    BIEMOND, J
    LAGENDIJK, RL
    MERSEREAU, RM
    [J]. PROCEEDINGS OF THE IEEE, 1990, 78 (05) : 856 - 883
  • [5] SPIRAL CT - DECREASED SPATIAL-RESOLUTION INVIVO DUE TO BROADENING OF SECTION-SENSITIVITY PROFILE
    BRINK, JA
    HEIKEN, JP
    BALFE, DM
    SAGEL, SS
    DICROCE, J
    VANNIER, MW
    [J]. RADIOLOGY, 1992, 185 (02) : 469 - 474
  • [6] CASTLEMAN KR, 1979, DIGITAL IMAGE PROCES, P139
  • [7] PULMONARY NODULE - EVALUATION WITH SPIRAL VOLUMETRIC CT
    COSTELLO, P
    ANDERSON, W
    BLUME, D
    [J]. RADIOLOGY, 1991, 179 (03) : 875 - 876
  • [8] COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION
    CRAWFORD, CR
    KING, KF
    [J]. MEDICAL PHYSICS, 1990, 17 (06) : 967 - 982
  • [9] SPIRAL CT OF THE PANCREAS
    DUPUY, DE
    COSTELLO, P
    ECKER, CP
    [J]. RADIOLOGY, 1992, 183 (03) : 815 - 818
  • [10] RESOLUTION ENHANCEMENT OF SPECTRA
    JANSSON, PA
    HUNT, RH
    PLYLER, EK
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1970, 60 (05) : 596 - &