CT SCANS THROUGH METAL SCANNING TECHNIQUE VERSUS HARDWARE COMPOSITION

被引:91
作者
HARAMATI, N
STARON, RB
MAZELSPERLING, K
FREEMAN, K
NICKOLOFF, EL
BARAX, C
FELDMAN, F
机构
[1] ALBERT EINSTEIN COLL MED,DEPT BIOSTAT,BRONX,NY 10467
[2] MONTEFIORE MED CTR,BRONX,NY 10467
[3] COLUMBIA PRESBYTERIAN MED CTR,DEPT RADIOL,NEW YORK,NY 10032
[4] MONTEFIORE MED CTR,BRONX,NY 10467
[5] BRONX ZOO,INT WILDLIFE CONSERVAT PK,BRONX,NY
[6] BRONX MUNICIPAL HOSP CTR,BRONX,NY 10461
[7] NEW YORK AQUARIUM,AQUARIAM WILDLIFE CONSERVAT,BROOKLYN,NY
[8] COLUMBIA UNIV COLL PHYS & SURG,NEW YORK,NY 10032
关键词
COMPUTED TOMOGRAPHY (CT) ARTIFACT; PROSTHESIS; HIP PROSTHESIS; METAL;
D O I
10.1016/0895-6111(94)90080-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Streak artifact on CT scans of metal containing areas has been a long standing problem. Although several artifact reducing methods have been used to improve image quality, most have been limited by requiring specialized equipment or lengthy complex calculations that are not automated. Others have shown that increasing the beam energy results in increased thickness of metal that may be imaged by CT without severe image degradation. We have studied the image quality of bone surrounding metal both with titanium and cobalt-chrome prostheses using various scanning techniques. Methods: In a double blind fashion, 28 radiology residents and attendings were surveyed as to the best technique for imaging bone detail surrounding metal. A series of images was arranged of an implanted titanium prosthesis, a cobalt-chrome prosthesis and a pelvis repaired with stainless steel pelvic reconstruction plates. Scans were performed using three techniques: 120 kVp, 170 mA, 2 s, 360-degrees rotation, 140 kVp, 140 mA, 3 s, 360-degrees rotation, 140 kVp, 140 mA, 4 s, 420-degrees rotation. Results: Titanium was superior to cobalt-chrome (p < .0001 Wilcoxon Signed Rank Test). No advantage was noted for higher kVp or increased scan arc of 420-degrees compared to the standard 360-degrees. Conclusion: Titanium allows improved bone detail surround the metal than CT cobalt-chrome. We have found no advantage to using either high kVp or a 420-degrees scan arc to improve the image quality of bone surrounded by metal.
引用
收藏
页码:429 / 434
页数:6
相关论文
共 19 条
  • [1] METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS
    AGINS, HJ
    ALCOCK, NW
    BANSAL, M
    SALVATI, EA
    WILSON, PD
    PELLICCI, PM
    BULLOUGH, PG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) : 347 - 356
  • [2] COMPUTED-TOMOGRAPHY IN THE ASSESSMENT AND PLANNING OF COMPLICATED TOTAL HIP-REPLACEMENT
    BARMEIR, E
    DUBOWITZ, B
    ROFFMAN, M
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1982, 53 (04): : 597 - 604
  • [3] INHIBITION OF APATITE FORMATION BY TITANIUM AND VANADIUM IONS
    BLUMENTHAL, NC
    COSMA, V
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH-APPLIED BIOMATERIALS, 1989, 23 (A1): : 13 - 22
  • [4] BRIEN WW, 1992, CLIN ORTHOP RELAT R, V276, P66
  • [5] Buly R L, 1992, J Arthroplasty, V7, P315, DOI 10.1016/0883-5403(92)90056-V
  • [6] REPROJECTION USING A PARALLEL BACKPROJECTOR
    CRAWFORD, CR
    [J]. MEDICAL PHYSICS, 1986, 13 (04) : 480 - 483
  • [7] EBRAHEIM NA, 1990, ORTHOPEDICS, V13, P1357
  • [8] THE POTENTIAL OF COMPUTED-TOMOGRAPHY IN VISUALIZING STRUCTURES INSIDE THE METAL CUP IN SURFACE-REPLACEMENT TOTAL HIP-ARTHROPLASTY
    EGUND, N
    PETTERSSON, H
    FROST, S
    LIDGREN, L
    [J]. SKELETAL RADIOLOGY, 1987, 16 (03) : 201 - 204
  • [9] COMPUTED-TOMOGRAPHY OF HUMERAL HEAD CUP ARTHROPLASTIES - A PRELIMINARY-REPORT
    EGUND, N
    JONSSON, E
    LIDGREN, L
    KELLY, I
    PETTERSSON, H
    [J]. ACTA RADIOLOGICA, 1987, 28 (01) : 71 - 73
  • [10] METALLIC HIP IMPLANTS - CT WITH MULTIPLANAR RECONSTRUCTION
    FISHMAN, EK
    MAGID, D
    ROBERTSON, DD
    BROOKER, AF
    WEISS, P
    SIEGELMAN, SS
    [J]. RADIOLOGY, 1986, 160 (03) : 675 - 681