SENSITIVE FLUOROMETRIC METHOD FOR THE DETERMINATION OF GLUTATHIONE AND SOME THIOLS IN BLOOD AND MAMMALIAN-TISSUES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY

被引:69
作者
TAKAHASHI, H
NARA, Y
MEGURO, H
TUZIMURA, K
机构
[1] Laboratory of Analytical Chemistry, Department of Food Chemistry, Tohoku University, Sendai, 980
来源
AGRICULTURAL AND BIOLOGICAL CHEMISTRY | 1979年 / 43卷 / 07期
关键词
D O I
10.1080/00021369.1979.10863636
中图分类号
S3 [农学(农艺学)];
学科分类号
0901 ;
摘要
A fluorescent reagent, N-(9-acridinyl)maleimide (NAM), was used for the determination of thiols in biological samples by high performance liquid chromatography* NAM-Iabeled glutathione (GSH), homocysteine, coenzyme A (CoA) and cysteine (CySH) were separated on a reversed-phase partition (octadecylsililated silica gel) column with the elution conditions of 0.06 m borate buffer pH 8.8: methanol (13: 1) at a flow rate of 0.6 ml/min within 15 min. In the absence of CoA in the sample, the elution conditions of 0.1 m borate buffer pH 8.8: methanol (15: 1) at a flow rate of 0.8 ml/min was used for the separations. Calibration curves were held up to 2.5 pmol for GSH and 11 pmol for CySH. About 0.17 µl of rat blood and 0.03 mg of rat liver equivalent to 0.1 nmol of GSH were determined. The sensitivity was 100 times higher than that obtained with an automatic amino acid analyzer. © 1979 Taylor and Francis Group, LLC.
引用
收藏
页码:1439 / 1445
页数:7