AN INTERFERENCE-FREE BIOSENSOR BASED ON GLUCOSE-OXIDASE ELECTROCHEMICALLY IMMOBILIZED IN A NONCONDUCTING POLY(PYRROLE) FILM FOR CONTINUOUS SUBCUTANEOUS MONITORING OF GLUCOSE THROUGH MICRODIALYSIS SAMPLING

被引:97
作者
PALMISANO, F
CENTONZE, D
GUERRIERI, A
ZAMBONIN, PG
机构
[1] UNIV BARI,DIPARTMENTO CHIM,CHIM ANALIT LAB,I-70126 BARI,ITALY
[2] UNIV BASILICATA,DIPARTIMENTO CHIM,CHIM ANALIT LAB,I-85100 POTENZA,ITALY
关键词
GLUCOSE SENSOR; MICRODIALYSIS; IN VIVO MONITORING; IMMOBILIZED ENZYMES; NONCONDUCTING POLYMERS;
D O I
10.1016/0956-5663(93)80023-I
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A glucose biosensor, based on glucose oxidase immobilized in a non-conducting (overoxidised) polypyrrole film, is described which proved practically immune from faradaic interference arising from endogeneous (ascorbate, urate, cysteine) and exogeneous (acetaminophen) electroactive interferents. The bias introduced in the measurement of 5 mM glucose by the given interferents at their maximum physiological levels never exceeded 2% which is, by far, the lowest value ever reported. The biosensor has been used for continuous subcutaneous monitoring of glucose in a rabbit implanted with a microdialysis probe. The potential and limits of this approach are discussed.
引用
收藏
页码:393 / 399
页数:7
相关论文
共 18 条
[1]   DESIGN AND INVITRO STUDIES OF A NEEDLE-TYPE GLUCOSE SENSOR FOR SUBCUTANEOUS MONITORING [J].
BINDRA, DS ;
ZHANG, YN ;
WILSON, GS ;
STERNBERG, R ;
THEVENOT, DR ;
MOATTI, D ;
REACH, G .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1692-1696
[2]   INTERFERENCE-FREE GLUCOSE SENSOR BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN AN OVEROXIDIZED NONCONDUCTING POLYPYRROLE FILM [J].
CENTONZE, D ;
GUERRIERI, A ;
MALITESTA, C ;
PALMISANO, F ;
ZAMBONIN, PG .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 342 (09) :729-733
[3]  
CENTONZE D, 1992, ANN CHIM-ROME, V82, P219
[4]   POTENTIALLY-IMPLANTABLE, FERROCENE-MEDIATED GLUCOSE SENSOR [J].
CLAREMONT, DJ ;
PENTON, C ;
PICKUP, JC .
JOURNAL OF BIOMEDICAL ENGINEERING, 1986, 8 (03) :272-274
[5]  
EGE H, 1989, Artificial Organs, V13, P171
[6]   INVIVO EVALUATION OF AN ELECTROENZYMATIC GLUCOSE SENSOR IMPLANTED IN SUBCUTANEOUS TISSUE [J].
JOHNSON, KW ;
MASTROTOTARO, JJ ;
HOWEY, DC ;
BRUNELLE, RL ;
BURDENBRADY, PL ;
BRYAN, NA ;
ANDREW, CC ;
ROWE, HM ;
ALLEN, DJ ;
NOFFKE, BW ;
MCMAHAN, WC ;
MORFF, RJ ;
LIPSON, D ;
NEVIN, RS .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (10) :709-714
[7]  
LUNTE CE, 1991, ANAL CHEM, V63, P773
[8]   GLUCOSE FAST-RESPONSE AMPEROMETRIC SENSOR BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN AN ELECTROPOLYMERIZED POLY(ORTHO-PHENYLENEDIAMINE) FILM [J].
MALITESTA, C ;
PALMISANO, F ;
TORSI, L ;
ZAMBONIN, PG .
ANALYTICAL CHEMISTRY, 1990, 62 (24) :2735-2740
[9]   INVIVO CONTINUOUS MONITORING OF GLUCOSE BY MICRODIALYSIS AND A GLUCOSE BIOSENSOR [J].
MASCINI, M ;
MOSCONE, D ;
BERNARDI, L .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 6 (1-3) :143-145
[10]   INVIVO CALIBRATION OF MICRODIALYSIS PROBES FOR EXOGENOUS COMPOUNDS [J].
MENACHERRY, S ;
HUBERT, W ;
JUSTICE, JB .
ANALYTICAL CHEMISTRY, 1992, 64 (06) :577-583