UNIVERSAL 4-PAIR IMPEDANCE BRIDGE

被引:16
作者
CORNEY, AC
机构
[1] Physics and Engineering Laboratory, Department of Scientific and Industrial Research, Lower Hutt
关键词
D O I
10.1109/TIM.1979.4314807
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A wide range 10 ppm four-terminal-pair audio-frequency impedance bridge is described. The bridge can be operated either in an impedance mode, for self-inductance, mutual inductance, and resistance measurements; or in an admittance mode for capacitance and conductance measurements. The frequency range for full accuracy is 40 Hz to 2 kHz. Maintenance of the calibration of the bridge is facilitated by the use of transformer ratios to extend the range of a small number of precision components, and extensive use is made of electronic techniques to automate auxiliary balances. Copyright © 1979 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:211 / 215
页数:5
相关论文
共 5 条
[1]   TECHNIQUES FOR COMPARING 4-TERMINAL-PAIR ADMITTANCE STANDARDS [J].
CUTKOSKY, RD .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION C-ENGINEERING AND INSTRUMENTATION, 1970, C 74 (3-4) :63-+
[2]  
KLASSEN NB, 1975, ELECTRON LETT, V11, P250
[3]   AC BRIDGE FOR PLATINUM-RESISTANCE THERMOMETRY [J].
THOMPSON, AM ;
SMALL, GW .
PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1971, 118 (11) :1662-&
[5]   STANDARDS FOR MEASUREMENT OF PHASE ANGLES OF RESISTORS [J].
THOMPSON, AM .
PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1975, 122 (05) :576-578