EVALUATION OF DIFFERENT TECHNIQUES FOR PEAK PURITY ASSESSMENT ON A DIODE-ARRAY DETECTOR IN LIQUID-CHROMATOGRAPHY

被引:29
作者
FABRE, H
LEBRIS, A
BLANCHIN, MD
机构
[1] Laboratoire de Chimie Analytique, Faculte ́de Pharmacie, 34060 Montpellier Ce ́dex 01
关键词
D O I
10.1016/0021-9673(94)00979-J
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The efficiencies of nine different techniques implemented on a diode-array detector in liquid chromatography for peak homogeneity determination were compared at a concentration level of a drug that can be found in biological samples. Solutions of cefotaxime sodium (analyte) spiked with various amounts of a model impurity (theophylline) were injected under chromatographic conditions giving severe overlap of the peaks (R(s)=0.14). Under the conditions used, the most efficient techniques implemented in the instrument were spectral suppression (ca. 0.5%), derivative spectrum (<1%), followed by spectral overlay, absorbance ratio plot (<5%), multiple-wavelength chromatograms overlay and purity parameter (ca. 5%). The numerical value of the absorbance ratio could also be exploited statistically to detect 0.5% of impurity. Theophylline at levels up to 10% could not be detected using three-dimensional plots, contour diagrams and derivative chromatograms.
引用
收藏
页码:81 / 88
页数:8
相关论文
共 10 条
[1]  
Poile, Conlon, J. Chromatogr., 204, (1981)
[2]  
Fell, Scott, Gill, Moffat, J. Chromatogr., 273, (1983)
[3]  
Clark, Fell, Scott, Wersterlund, J. Chromatogr., 286, (1984)
[4]  
Fell, Clark, Scott, J. Chromatogr., 297, (1984)
[5]  
Alfredson, Sheehan, Lenert, Aamodt, Correia, J. Chromatogr., 385, (1987)
[6]  
Chan, Carr, J. Pharm. Biomed., Anal., 8, (1990)
[7]  
Fabre, Fell, J. Liq. Chromatogr., 15, (1992)
[8]  
Pharmacope´e Europe´enne, (1987)
[9]  
Owino, Clark, Fell, J. Chromatogr. Sci., 29, (1991)
[10]  
Product Note, (1993)