A TIME DERANDOMIZER FOR ANALOG PULSES

被引:2
作者
ROBINSON, LB
GIN, F
机构
[1] Lawrence Radiation Laboratory, University of California, Berkeley, CA
来源
NUCLEAR INSTRUMENTS & METHODS | 1969年 / 75卷 / 01期
关键词
D O I
10.1016/0029-554X(69)90663-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A storage unit for analog pulses is described which is used to decrease counting losses in high-rate, high-resolution, pulse-height analysis experiments. Three pulse stretchers connected in series store up to three pulses until they can be processed by an external analog-to-digital converter. An effective dead time of 6 μs is achieved when analyzing randomly occuring pulses at rates up to 20 000 cps. No degradation in pulse height resolution due to the insertion of this instrument has been detected. © 1969.