A ROENTGENOGRAPHIC METHOD FOR RAPID EVALUATION OF X-RAY MACHINES

被引:2
作者
WING, K
SOREMARK, R
SAIRENJI, E
机构
[1] Harvard School of Dental Medicine Boston, MA
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS | 1968年 / 25卷 / 06期
关键词
D O I
10.1016/0030-4220(68)90154-0
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
A new method in which Polaroid type 52 Land films are used for the evaluation of x-ray machines is presented. The black and white print, which is developed in ten seconds, demonstrates the adequacy of collimation, the aiming of the central beam, and any scattering or leakage of radiation in the anterior direction. By comparison with standard exposures, an accurate estimate can be made of the roentgen output of the x-ray machine. The method is very rapid and provides a more than adequate substitute for the fluorescent screen and radiation recorders now in use for direct evaluations of x-ray machines. © 1968.
引用
收藏
页码:822 / &
相关论文
共 13 条
[2]   A MEDICAL X-RAY SURVEY PACKET [J].
DITCHEK, T ;
FRISOLI, A ;
JANES, D .
HEALTH PHYSICS, 1966, 12 (03) :341-+
[3]  
HOLLENDER L, 1962, ODONT REVY, V13, P34
[4]  
MILLER J W, 1965, Radiol Health Data Rep, V6, P271
[5]  
MILLER JW, 1963, RADIOL HEALTH DATA, V4, P41
[6]  
PRATHER P F, 1964, Radiol Health Data Rep, V5, P341
[7]  
SENN C L, 1963, Radiol Health Data Rep, V4, P215
[8]   A FILM METHOD FOR ESTIMATING PEAK KILOVOLTAGE TOTAL FILTRATION AND ROENTGEN OUTPUT OF SELF-RECTIFIED DENTAL X-RAY MACHINES [J].
WINKLER, KG ;
LEVIN, SG .
HEALTH PHYSICS, 1966, 12 (03) :345-+
[9]  
1960, 1 FED RAD COUNC PROT
[10]  
1961, 76 NAT BUR STAND HAN