OPTIMIZATION OF DETECTION SENSITIVITY IN THE CAPILLARY ELECTROPHORESIS OF INORGANIC ANIONS

被引:252
作者
JANDIK, P
JONES, WR
机构
[1] Ions Analysis Department, Waters Chromatography Division, Millipore Corporation, Milliford, MA 01757
来源
JOURNAL OF CHROMATOGRAPHY | 1991年 / 546卷 / 1-2期
关键词
D O I
10.1016/S0021-9673(01)93042-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Published reports describing applications of various capillary electrophoresis methods to the analysis of low-molecular-weight anionic species are reviewed and evaluated. An attempt is made to summarize the parameters relevant to the optimization of detection sensitivity in capillary electrophoresis. Based on a comparison with the results obtained by other techniques, indirect photometry is shown to have a number of important advantages for the detection of anions. For a given carrier electrolyte, the range of linearity can be extended by increasing the concentration of the anion responsible for background UV absorption. With chromate-containing carrier electrolytes the linearity range could be extended over more than three orders of magnitude. Good correlation between the relative molar absorptivities and sensitivity was found for six aromatic carboxylic acids serving as carrier electrolytes. Using optimized conditions for electromigrative sample introduction, the achievable detection limits are found to be in the low nanomolar range. This represents more than a 100-fold increase in sensitivity over the results obtained by hydrodynamic sample introduction. An isotachophoretic steady state occurring during the electromigration is presented as an explanation for the trace enrichment of low levels of analytes during sample introduction by electromigration.
引用
收藏
页码:431 / 443
页数:13
相关论文
共 19 条
[1]   DETERMINATION OF METAL-ION COMPLEXES IN ELECTROPLATING SOLUTIONS USING CAPILLARY ZONE ELECTROPHORESIS WITH UV DETECTION [J].
AGUILAR, M ;
HUANG, XH ;
ZARE, RN .
JOURNAL OF CHROMATOGRAPHY, 1989, 480 :427-431
[2]   USE OF A DOUBLE-DETECTOR SYSTEM FOR THE MEASUREMENT OF MOBILITIES IN ZONE ELECTROPHORESIS [J].
BECKERS, JL ;
VERHEGGEN, TPEM ;
EVERAERTS, FM .
JOURNAL OF CHROMATOGRAPHY, 1988, 452 :591-600
[3]   INDIRECT PHOTOMETRIC DETECTION IN CAPILLARY ZONE ELECTROPHORESIS [J].
FORET, F ;
FANALI, S ;
OSSICINI, L ;
BOCEK, P .
JOURNAL OF CHROMATOGRAPHY, 1989, 470 (01) :299-308
[4]   ONLINE FIBER OPTIC UV DETECTION CELL AND CONDUCTIVITY CELL FOR CAPILLARY ZONE ELECTROPHORESIS [J].
FORET, F ;
DEML, M ;
KAHLE, V ;
BOCEK, P .
ELECTROPHORESIS, 1986, 7 (09) :430-432
[5]  
Foret F, 1990, J MICROCOLUMN SEP, V2, P229
[6]   DETERMINATION OF NITRATE, CHLORIDE AND SULFATE IN DRINKING-WATER BY CAPILLARY FREE-ZONE ELECTROPHORESIS [J].
GEBAUER, P ;
DEML, M ;
BOCEK, P ;
JANAK, J .
JOURNAL OF CHROMATOGRAPHY, 1983, 267 (02) :455-457
[7]  
HADDAD PR, 1990, J CHROMATOGRAPHY LIB, V46
[8]  
Hjerten S, 1967, Chromatogr Rev, V9, P122, DOI 10.1016/0009-5907(67)80003-6
[9]   CARRIER-FREE ZONE ELECTROPHORESIS, DISPLACEMENT ELECTROPHORESIS AND ISOELECTRIC-FOCUSING IN A HIGH-PERFORMANCE ELECTROPHORESIS APPARATUS [J].
HJERTEN, S ;
ELENBRING, K ;
KILAR, F ;
LIAO, JL ;
CHEN, AJC ;
SIEBERT, CJ ;
ZHU, MD .
JOURNAL OF CHROMATOGRAPHY, 1987, 403 :47-61
[10]   EFFECT OF ELECTROLYTE AND SAMPLE CONCENTRATION ON THE RELATIONSHIP BETWEEN SENSITIVITY AND RESOLUTION IN CAPILLARY ZONE ELECTROPHORESIS USING CONDUCTIVITY DETECTION [J].
HUANG, XH ;
GORDON, MJ ;
ZARE, RN .
JOURNAL OF CHROMATOGRAPHY, 1989, 480 :285-288