THE DINAMAP-1846SX AUTOMATED BLOOD-PRESSURE RECORDER - COMPARISON WITH THE HAWKSLEY RANDOM ZERO SPHYGMOMANOMETER UNDER FIELD CONDITIONS

被引:62
作者
WHINCUP, PH [1 ]
BRUCE, NG [1 ]
COOK, DG [1 ]
SHAPER, AG [1 ]
机构
[1] UNIV LONDON, SCH MED, LONDON NW3 2PF, ENGLAND
基金
英国惠康基金;
关键词
D O I
10.1136/jech.46.2.164
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Study objective - The aim was to compare the performance of the Dinamap 1846SX automated oscillometric blood pressure recorder with that of the Hawksley random zero sphygmomanometer during use under field study conditions. Design - Two independent within subject measurement comparisons were made, one in adults and one in children, each conducted in three stages over several months while the Dinamap instruments were being used in epidemiological field surveys. Setting - The studies were done in outpatients clinics (adults) and primary schools (children). Participants - 141 adults (20-85 years) and 152 children (5-7 years) took part. Measurements and main results - In adults a pair of measurements was made with each instrument, the order alternating for consecutive subjects. In children one measurements was made with each instrument, in random order. Measurements with the Dinamap 1846SX were higher than those with the random zero sphygmomanometer both in adults (mean difference 8.1 mm Hg; 95% CI 6.5 to 9.7 mm Hg) and in children (mean difference 8.3 mm Hg; 95% CI 6.9 to 9.7 mm Hg). Diastolic measurements were on average very similar both in adults and in children. The results were consistent at all three stages of both studies. The differences in systolic measurement were independent of blood pressure level. However, the extent of agreement in diastolic pressure depended on the diastolic blood pressure level; in both studies Dinamap diastolic measurements were higher at low diastolic pressures while random zero diastolic measurements were higher at high diastolic pressures. Conclusions - Systolic measurements made with the Dinamap 1846SX instrument are not directly comparable with those of the Hawksley random zero sphygmomanometer and are unlikely to be comparable with those of earlier Dinamap models. These differences have important implications for clinical practice and for comparisons of blood pressure measurement between epidemiological studies. However, the consistency of measurement by the Dinamap 1846SX over time suggests that the instrument may have a place in standardised blood pressure measurement in the research setting.
引用
收藏
页码:164 / 169
页数:6
相关论文
共 36 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   NON-INVASIVE ESTIMATION OF CENTRAL AORTIC PRESSURE USING THE OSCILLOMETRIC METHOD FOR ANALYZING SYSTEMIC ARTERY PULSATILE BLOOD-FLOW - COMPARATIVE-STUDY OF INDIRECT SYSTOLIC, DIASTOLIC, AND MEAN BRACHIAL-ARTERY PRESSURE WITH SIMULTANEOUS DIRECT ASCENDING AORTIC PRESSURE MEASUREMENTS [J].
BOROW, KM ;
NEWBURGER, JW .
AMERICAN HEART JOURNAL, 1982, 103 (05) :879-886
[3]  
BRUCE NG, 1988, J HYPERTENS, V6, P375
[4]   GEOGRAPHICAL VARIATIONS IN BLOOD-PRESSURE IN BRITISH MEN AND WOMEN [J].
BRUCE, NG ;
COOK, DG ;
SHAPER, AG ;
THOMSON, AG .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 1990, 43 (04) :385-398
[5]   NON-INVASIVE DETERMINATION OF SYSTOLIC, DIASTOLIC AND END-SYSTOLIC BLOOD-PRESSURE IN NEONATES, INFANTS AND YOUNG-CHILDREN - COMPARISON WITH CENTRAL AORTIC PRESSURE MEASUREMENTS [J].
COLAN, SD ;
FUJII, A ;
BOROW, KM ;
MACPHERSON, D ;
SANDERS, SP .
AMERICAN JOURNAL OF CARDIOLOGY, 1983, 52 (07) :867-870
[6]   THE RANDOM-ZERO VERSUS THE STANDARD MERCURY SPHYGMOMANOMETER - A SYSTEMATIC BLOOD-PRESSURE DIFFERENCE [J].
DEGAUDEMARIS, R ;
FOLSOM, AR ;
PRINEAS, RJ ;
LUEPKER, RV .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 1985, 121 (02) :282-290
[7]   A DEVICE FOR THE AUTOMATIC-MEASUREMENT OF BLOOD-PRESSURE IN EPIDEMIOLOGIC STUDIES [J].
ELLISON, RC ;
GAMBLE, WJ ;
TAFT, DS .
AMERICAN JOURNAL OF EPIDEMIOLOGY, 1984, 120 (04) :542-549
[8]  
EVANS JG, 1970, BRIT J PREV SOC MED, V24, P10
[9]  
FRIESEN RH, 1981, ANESTH ANALG, V60, P742
[10]  
FROHLICH ED, 1988, CIRCULATION, V77, pA502