EVALUATION OF THE STERIS-SYSTEM-1 PERACETIC-ACID ENDOSCOPE PROCESSOR

被引:36
作者
BRADLEY, CR
BABB, JR
AYLIFFE, GAJ
机构
[1] Hospital Infection Research Laboratory, City Hospital NHS Trust, Birmingham, BIS 7QH, Dudley Road
关键词
ENDOSCOPE DECONTAMINATION; PERACETIC ACID;
D O I
10.1016/0195-6701(95)90196-5
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
An automated endoscope sterilizing machine, the Steris System 1 Processor, was tested for bactericidal and sporicidal efficacy. The disinfectant, peracetic acid, was diluted to 0.2% within an enclosed system. The exposure time to the disinfectant was 12 min and the overall cycle time ranged from 25-38 min, mean 29 min. Preliminary suspension tests, with and without yeast or serum, showed a log(10) reduction of > 5 with Pseudomonas aeruginosa, Staphylococcus aureus and Bacillus subtilis in 5 min with 0.2% peracetic acid. After a routine cycle in the machine, endoscopes contaminated with the same organisms showed no growth. Two of 24 spore strips, containing 10(6) B. subtilis showed a small number of survivors (less than 10 per strip). No significant damage to the endoscope was observed although the number of cycles tested was small (i.e. 31). The advantage of the system is that staff are not directly exposed to the agent, but the costs per cycle are higher than glutaraldehyde, since peracetic acid is not renewed. Unlike other automated processers the Steris machine has no cleaning cycle.
引用
收藏
页码:143 / 151
页数:9
相关论文
共 12 条
  • [1] STERILIZATION OF ARTHROSCOPES AND LAPAROSCOPES
    AYLIFFE, GAJ
    BABB, JR
    BRADLEY, CR
    [J]. JOURNAL OF HOSPITAL INFECTION, 1992, 22 (04) : 265 - 269
  • [2] Babb J R, 1980, J Hosp Infect, V1, P63, DOI 10.1016/0195-6701(80)90033-X
  • [3] Babb J R, 1992, J Hosp Infect, V20, P51, DOI 10.1016/0195-6701(92)90061-P
  • [4] DISINFECTION AND STERILIZATION OF ENDOSCOPES
    BABB, JR
    [J]. CURRENT OPINION IN INFECTIOUS DISEASES, 1993, 6 (04) : 532 - 537
  • [5] Block S. S., 1991, DISINFECTION STERILI, P167
  • [6] BRADLEY CR, 1994, NURS TIMES, V13, P70
  • [7] BURGE PS, 1989, BRIT MED J, V300, P299
  • [8] DECONTAMINATION OF UROLOGICAL EQUIPMENT - INTERIM-REPORT OF A WORKING GROUP OF THE STANDING COMMITTEE ON UROLOGICAL INSTRUMENTS OF THE BRITISH-ASSOCIATION-OF-UROLOGICAL-SURGEONS
    COOKE, RPD
    FENELEY, RCL
    AYLIFFE, G
    LAWRENCE, WT
    EMMERSON, AM
    GREENGRASS, SM
    [J]. BRITISH JOURNAL OF UROLOGY, 1993, 71 (01): : 5 - 9
  • [9] ALDEHYDE DISINFECTANTS AND HEALTH IN ENDOSCOPY UNITS - THE REPORT OF A WORKING PARTY OF THE BRITISH SOCIETY OF GASTROENTEROLOGY ENDOSCOPY COMMITTEE
    COWAN, RE
    MANNING, AP
    AYLIFFE, GAJ
    AXON, ATR
    CAUSTON, JS
    CRIPPS, NF
    HALL, R
    HANSON, PJV
    HARRISON, J
    LEICESTER, RJ
    NEUMANN, C
    WICKS, J
    [J]. GUT, 1993, 34 (11) : 1641 - 1645
  • [10] CROW S, 1992, INFECT CONT HOSP EP, V13, P111