A SIMPLE CALIBRATION METHOD FOR MECHANICALLY BRAKED CYCLE ERGOMETERS

被引:38
作者
VANPRAAGH, E
BEDU, M
RODDIER, P
COUDERT, J
机构
[1] UFR SCI AUBIERE,DEPT MATH,F-63172 CLERMONT FERRAND,FRANCE
[2] FAC MED CLERMONT,PHYSIOL LAB,F-63001 CLERMONT FERRAND,FRANCE
关键词
CYCLE ERGOMETER; CALIBRATION; MECHANICAL POWER; EXERCISE PHYSIOLOGY;
D O I
10.1055/s-2007-1021229
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
The calibration of cycle ergometers should be checked regularly. Some studies have showen calibration errors of more than 40%. A simple, inexpensive calibrating method for mechanically braked cycle ergometers was developed and tried out on a new type of ergocycle. The cycle ergometer was elevated and the crank replaced by a pulley fitted to the shaft. The crank speed (rpm) increased linearly as a function of time when different masses were applied on the pulley. For a given braking force on the cycle ergometer, different accelerations corresponding to the increased pulley forces could be measured. When extrapolating for zero acceleration, it was possible to determine a "limit-force" which allowed the system to be in equilibrium. Additional force creates motion. The same experiments were repeated with increasing braking forces. Using the differently sized gear sprockets of the transmission system, it was possible to calculate the actual force, including all the resistances. The actual force found by the calibrating method was then compared with the indicated force proposed by the manufacturer. With increasing forces, the relative errors decreased from 9.6 to 2.9 %. The cycle ergometer calibrated by this technique meets the standards recommended in exercise physiology.
引用
收藏
页码:27 / 30
页数:4
相关论文
共 14 条
[1]   THE WINGATE ANAEROBIC TEST - AN UPDATE ON METHODOLOGY, RELIABILITY AND VALIDITY [J].
BAROR, O .
SPORTS MEDICINE, 1987, 4 (06) :381-394
[2]  
BEDU M, 1991, J APPL PHYSIOL, V70, P1031, DOI 10.1152/jappl.1991.70.3.1031
[3]  
BRUE F, 1983, 5TH P M RSG BRUSS
[4]   BIOLOGICAL AND TECHNOLOGICAL VARIABILITY OF 3 ANAEROBIC ERGOMETER TESTS [J].
COGGAN, AR ;
COSTILL, DL .
INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1984, 5 (03) :142-145
[5]  
HOFFMANN K, 1983, STANDARDISIERUNG KAL, P14
[6]  
Mellerowicz H., 1983, STANDARDISIERUNG KAL
[7]  
MELLEROWICZ H, 1981, ERGOMETRY BASIC MED, P45
[8]   DYNAMIC TORQUEMETER CALIBRATION OF BICYCLE ERGOMETERS [J].
RUSSELL, JC ;
DALE, JD .
JOURNAL OF APPLIED PHYSIOLOGY, 1986, 61 (03) :1217-1220
[9]  
RYBCZYNSKI N, 1987, THESIS MAINZ
[10]  
SMODLAKA VN, 1981, ERGOMETRY BASICS EXE, P385