NEW APPARATUS FOR MEASURING RADON ADSORPTION ON SOLID ADSORBENTS

被引:8
作者
HASSAN, NM
HINES, AL
GHOSH, TK
LOYALKA, SK
KETRING, AR
机构
[1] UNIV MISSOURI, DEPT CHEM ENGN, NUCL ENGN & PARTICULATE SYST RES CTR, COLUMBIA, MO 65211 USA
[2] UNIV MISSOURI, RES REACTOR, COLUMBIA, MO 65211 USA
关键词
D O I
10.1021/ie00057a024
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A new experimental system was designed to measure radon uptake by solid adsorbents from air or other carrier gases/vapors. The total amount of radon adsorbed corresponding to a specific gas-phase concentration was determined by simultaneously measuring the solid-phase and gas-phase concentrations. The system was used to measure radon adsorption isotherms on BPL activated carbon at 288, 298, and 308 K and on silica gel and molecular sieve 13X at 298 K. The isotherms were of type III according to Brunauer's classification. The heat of adsorption data indicated that the BPL activated carbon provided a heterogeneous surface for radon adsorption. The equilibrium data were correlated by the Freundlich equation. The possible adsorption mechanism and the use of the adsorption isotherms to measure indoor radon concentrations are discussed.
引用
收藏
页码:2205 / 2211
页数:7
相关论文
共 18 条
[1]  
ACKLEY RD, 1975, ORNLTM4883 CHEM TECH
[2]  
ADAMSON AW, 1976, PHYSICAL CHEM SURFAC, P389
[3]   Adsorption of gases in multimolecular layers [J].
Brunauer, S ;
Emmett, PH ;
Teller, E .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1938, 60 :309-319
[4]   MIXED ADSORPTION OF RADON AND ARGON ON SILICA GEL [J].
BURTT, BP ;
KURBATOV, JD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1948, 70 (06) :2278-2282
[5]   A DIFFUSION BARRIER CHARCOAL ADSORPTION COLLECTOR FOR MEASURING RN CONCENTRATIONS IN INDOOR AIR [J].
COHEN, BL ;
NASON, R .
HEALTH PHYSICS, 1986, 50 (04) :457-463
[6]   THEORY AND PRACTICE OF RADON MONITORING WITH CHARCOAL ADSORPTION [J].
COHEN, BL ;
COHEN, ES .
HEALTH PHYSICS, 1983, 45 (02) :501-508
[7]  
COLEMAN RD, 1957, AM IND HYG ASSOC Q, P17
[8]  
Freundlich H, 1906, Z PHYS CHEM-STOCH VE, V57, P385
[9]   PASSIVE, INTEGRATED MEASUREMENT OF INDOOR RADON USING ACTIVATED CARBON [J].
GEORGE, AC .
HEALTH PHYSICS, 1984, 46 (04) :867-872
[10]  
GUBELI VO, 1954, HELV CHIM ACTA, V37, P22224