CAPILLARY ISOELECTRIC-FOCUSING ELECTROSPRAY MASS-SPECTROMETRY FOR PROTEIN-ANALYSIS

被引:100
作者
TANG, Q
HARRATA, AK
LEE, CS
机构
[1] IOWA STATE UNIV SCI & TECHNOL,DEPT CHEM,AMES,IA 50011
[2] IOWA STATE UNIV SCI & TECHNOL,AMES LAB,AMES,IA 50011
关键词
D O I
10.1021/ac00115a021
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
On-line combination of capillary isoelectric focusing (CIEF) with electrospray mass spectrometry (ESMS) as a two-dimensional separation system is demonstrated. Mixtures of model proteins including cytochrome c (horse heart), myoglobin (horse heart), and carbonic anhydrase II (bovine erythrocyte) are focused and cathodically mobilized in a polyacrylamide-coated capillary. At the end of CIEF capillary, the mobilized protein zones are analyzed by mass spectrometry coupled on-line to an electrospray interface with a coaxial sheath now configuration. The effects of carrier ampholyte concentration on the CIEF separation and the protein electrospray ionization mass spectra are presented and discussed. In this study, the focusing effect of CIEF permits analysis of very dilute protein samples. A typical concentration factor of 50-100 times is observed. The concentration detection limit of myoglobin for a full-scan CIEF-ESMS analysis is in the range of 10(-7) M, 2 orders of magnitude over that possible with normal capillary zone electrophoresis ESMS.
引用
收藏
页码:3515 / 3519
页数:5
相关论文
共 34 条
[1]  
ASHTON DS, 1993, ORG MASS SPECTROM, V28, P579
[2]   ON-COLUMN SAMPLE CONCENTRATION USING FIELD AMPLIFICATION IN CZE [J].
CHIEN, RL ;
BURGI, DS .
ANALYTICAL CHEMISTRY, 1992, 64 (08) :A489-A496
[3]   PROBING CONFORMATIONAL-CHANGES IN PROTEINS BY MASS-SPECTROMETRY [J].
CHOWDHURY, SK ;
KATTA, V ;
CHAIT, BT .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1990, 112 (24) :9012-9013
[4]  
CREIGHTON TE, 1990, PROTEIN STRUCTURE PR, pCH3
[5]   CAPILLARY ELECTROPHORESIS ELECTROSPRAY IONIZATION-MASS SPECTROMETRY [J].
EDMONDS, CG ;
LOO, JA ;
BARINAGA, CJ ;
UDSETH, HR ;
SMITH, RD .
JOURNAL OF CHROMATOGRAPHY, 1989, 474 (01) :21-37
[6]   ONLINE TIME-OF-FLIGHT MASS-SPECTROMETRIC ANALYSIS OF PEPTIDES SEPARATED BY CAPILLARY ELECTROPHORESIS [J].
FANG, LL ;
ZHANG, R ;
WILLIAMS, ER ;
ZARE, RN .
ANALYTICAL CHEMISTRY, 1994, 66 (21) :3696-3701
[7]   ELECTROSPRAY IONIZATION-PRINCIPLES AND PRACTICE [J].
FENN, JB ;
MANN, M ;
MENG, CK ;
WONG, SF ;
WHITEHOUSE, CM .
MASS SPECTROMETRY REVIEWS, 1990, 9 (01) :37-70
[8]   ELECTROSPRAY IONIZATION FOR MASS-SPECTROMETRY OF LARGE BIOMOLECULES [J].
FENN, JB ;
MANN, M ;
MENG, CK ;
WONG, SF ;
WHITEHOUSE, CM .
SCIENCE, 1989, 246 (4926) :64-71
[9]   ION FORMATION FROM CHARGED DROPLETS - ROLES OF GEOMETRY, ENERGY, AND TIME [J].
FENN, JB .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1993, 4 (07) :524-535
[10]   ON-COLUMN TRANSIENT AND COUPLED COLUMN ISOTACHOPHORETIC PRECONCENTRATION OF PROTEIN SAMPLES IN CAPILLARY ZONE ELECTROPHORESIS [J].
FORET, F ;
SZOKO, E ;
KARGER, BL .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :3-12