SPURIOUS DISCHARGES DUE TO LATE INSULATION BREAK IN ENDOCARDIAL SENSING LEADS FOR CARDIOVERTER-DEFIBRILLATORS

被引:11
作者
PETERS, RW
FOSTER, AH
SHOROFSKY, SR
NOLAN, DA
GOLD, MR
机构
[1] UNIV MARYLAND HOSP,DEPT MED,DIV CARDIOL,BALTIMORE,MD 21201
[2] UNIV MARYLAND HOSP,DEPT CARDIOTHORAC SURG,BALTIMORE,MD 21201
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1995年 / 18卷 / 03期
关键词
CARDIOVERTER DEFIBRILLATOR; ENDOCARDIAL LEADS; INSULATION BREAK;
D O I
10.1111/j.1540-8159.1995.tb02549.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Despite their similarity to permanent pacemaker leads, endocardial sensing leads for cardioverter defibrillators have a relatively high failure rate. We describe four patients with endocardial rate sensing leads who developed inappropriate discharges 10-30 months after implantation due to small breaks in the lead insulation. This problem may become increasingly common as the number of cardioverter defibrillator implants with transvenous leads continues to grow and should be considered in the differential diagnosis of late sensing failure or inappropriate device discharges.
引用
收藏
页码:478 / 481
页数:4
相关论文
共 7 条
  • [1] LONG-TERM COMPLICATIONS OF IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LEAD SYSTEMS
    ALMASSI, GH
    OLINGER, GN
    WETHERBEE, JN
    FEHL, G
    [J]. ANNALS OF THORACIC SURGERY, 1993, 55 (04) : 888 - 892
  • [2] SURVIVAL OF IMPLANTABLE PACEMAKER LEADS
    FURMAN, S
    BENEDEK, ZM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1910 - 1914
  • [3] ELECTROCARDIOGRAPHICALLY DOCUMENTED UNNECESSARY, SPONTANEOUS SHOCKS IN 241 PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS
    GRIMM, W
    FLORES, BF
    MARCHLINSKI, FE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11): : 1667 - 1673
  • [4] ONUNAIN SS, 1993, J AM COLL CARDIOL, V21, pA303
  • [5] LONG-TERM MULTICENTER EXPERIENCE WITH A 2ND-GENERATION IMPLANTABLE PACEMAKER-DEFIBRILLATOR IN PATIENTS WITH MALIGNANT VENTRICULAR TACHYARRHYTHMIAS
    SAKSENA, S
    POCZOBUTTJOHANOS, M
    CASTLE, LW
    FOGOROS, RN
    ALPERT, BL
    KRON, J
    PACIFICO, A
    GRIFFIN, J
    RUSKIN, JN
    KEHOE, RF
    YEE, R
    DORIAN, P
    KERR, CR
    LUCERI, RM
    POLISENO, M
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (03) : 490 - 499
  • [6] SENSING PACING LEAD COMPLICATIONS WITH A NEWER GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - WORLDWIDE EXPERIENCE FROM THE GUARDIAN-ATP-4210 CLINICAL-TRIAL
    STAMBLER, BS
    WOOD, MA
    DAMIANO, RJ
    GREENWAY, PS
    SMUTKA, ML
    ELLENBOGEN, KA
    BATSFORD, W
    BAUERNFEIND, R
    BECKMAN, K
    BROWNSTEIN, S
    CASTLE, L
    DEERING, T
    ELLENBOGEN, K
    ESTES, NAM
    FRANKLIN, J
    FREIHLING, T
    HAISTY, K
    KEHOE, R
    LOVE, C
    LUCERI, R
    NACCARELLI, G
    PACIFICO, A
    PARSONNET, V
    PETROPOULOS, T
    SAKSENA, S
    SHARMA, A
    SINGER, I
    VENDITTI, F
    WALTER, P
    ZIPES, D
    DAVIS, MJ
    HELLESTRAND, K
    KUCHAR, D
    POWELL, A
    STAFFORD, W
    UTHER, J
    VOHRA, J
    DORIAN, P
    GULAMHUSEIN, S
    KERR, CR
    ROSENGARTEN, M
    ROY, D
    TANG, ASL
    WILLIAMS, W
    YEE, R
    MOLLER, M
    THOMSEN, PEB
    ALIOT, E
    LECLERQ, JF
    LEMAREC, H
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (01) : 123 - 132
  • [7] LONG-TERM OUTCOME WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    WINKLE, RA
    MEAD, RH
    RUDER, MA
    GAUDIANI, VA
    SMITH, NA
    BUCH, WS
    SCHMIDT, P
    SHIPMAN, T
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (06) : 1353 - 1361