A SIMPLE PRECISION FLUXMETER

被引:3
作者
SASAKI, H
机构
[1] Institute for Nuclear Study, University of Tokyo, Tanashi-shi, Tokyo
来源
NUCLEAR INSTRUMENTS & METHODS | 1969年 / 76卷 / 01期
关键词
D O I
10.1016/0029-554X(69)90296-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A flip-coil fluxmeter coupled with an integrating digital voltmeter was developed. With small modification of the digital voltmeter, high precision with the sensitivity of several milligauss was obtained without any elaborate mechanical device for flipping the search coil. Measuring accuracy as good as a few parts in 105 can be achieved at the hundred gauss region with simple treatment of the fluxmeter. © 1969.
引用
收藏
页码:100 / &
相关论文
共 7 条
[1]  
BERVILL EC, 1967, P INT C MAGNET TECHN, P675
[2]  
BUSH GE, 1965, P INT S MAGNET TECHN, P407
[3]   BIBLIOGRAPHICAL REVIEW OF THE METHODS OF MEASURING MAGNETIC FIELDS [J].
GERMAIN, C .
NUCLEAR INSTRUMENTS & METHODS, 1963, 21 (01) :17-46
[4]  
GRAY DA, 1963, NIRLR4
[5]  
HIROSE T, 1969, SJCA693
[6]   EQUIPMENT + PRECEDURES USED IN MAGNETIC FIELD MEASUREMENTS ON MURA 50-MEV ELECTRON ACCELERATOR. 7 [J].
MULADY, JR ;
HAXBY, RO ;
PRUETT, CH ;
WALLENMEYER, WA ;
BUSH, GE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1964, 35 (11) :1437-&
[7]  
WATSON PG, 1965, P INT S MAGNET TECHN, P393