STUDY OF A GLASSY-CARBON ELECTRODE IN AMPEROMETRIC DETECTION USING DC, NORMAL AND DIFFERENTIAL PULSE TECHNIQUES

被引:30
作者
DIEKER, JW
VANDERLINDEN, WE
POPPE, H
机构
[1] Laboratory for Analytical Chemistry, University of Amsterdam, 1018 WV Amsterdam
关键词
D O I
10.1016/0039-9140(79)80081-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The results of a comparative study on d.c., normal pulse and differential pulse techniques applied to anodic amperometric detection at a glassy carbon electrode in a voltammetric flow-through cell are presented. The important aspects examined are response time, linearity, limit of detection and selectivity. It is shown that the d.c. mode is the most favourable as long as no adsorption of oxidation products takes place. If strong adsorption occurs, normal pulse detection is recommended, although the limit of detection is somewhat larger. © 1979.
引用
收藏
页码:511 / 518
页数:8