AN IMPROVED INVIVO NEUTRON-ACTIVATION SYSTEM FOR MEASURING KIDNEY CADMIUM

被引:13
作者
FRANKLIN, DM [1 ]
ARMSTRONG, R [1 ]
CHETTLE, DR [1 ]
SCOTT, MC [1 ]
机构
[1] UNIV BIRMINGHAM,SCH PHYS & SPACE RES,BIRMINGHAM B15 2TT,W MIDLANDS,ENGLAND
关键词
D O I
10.1088/0031-9155/35/10/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
An in vivo neutron activation system for measuring kidney cadmium has been redesigned, firstly to reduce ambient dose levels and, secondly, to improve the cadmium signal to neutron dose ratio by modifying the neutron spectrum from 238Pu/Be, by interposing a beryllium premoderator. The ambient dose was reduced by a factor of seven. The overall system performance (lower limit of detection for a given dose) was improved by 40-50%. This 238Pu/Be based system performs as well as or better than analogous 252Cf systems, without the drawback of the relatively short half life of 252Cf.
引用
收藏
页码:1397 / 1408
页数:12
相关论文
共 11 条
[1]
ALHADDAD IK, 1981, INT J APPL RADIAT IS, V32, P109, DOI 10.1016/0020-708X(81)90143-5
[2]
INVIVO AND INVITRO MEASUREMENTS OF LEAD AND CADMIUM [J].
CHETTLE, DR ;
FRANKLIN, DM ;
GUTHRIE, CJG ;
SCOTT, MC ;
SOMERVAILLE, LJ .
BIOLOGICAL TRACE ELEMENT RESEARCH, 1987, 13 :191-208
[3]
CADMIUM FUME INHALATION AND EMPHYSEMA [J].
DAVISON, AG ;
TAYLOR, AJN ;
DARBYSHIRE, J ;
CHETTLE, DR ;
GUTHRIE, CJG ;
OMALLEY, D ;
MASON, HJ ;
FAYERS, PM ;
VENABLES, KM ;
PICKERING, CAC ;
FRANKLIN, D ;
SCOTT, MC ;
HOLDEN, H ;
WRIGHT, AL ;
GOMPERTZ, D .
LANCET, 1988, 1 (8587) :663-667
[4]
ELLIS KJ, 1983, NEUROTOXICOLOGY, V4, P164
[5]
MASON HJ, 1988, BRIT J IND MED, V45, P793
[6]
CADMIUM - INVIVO DETECTION IN MAN [J].
MCLELLAN, JS ;
THOMAS, BJ ;
FREMLIN, JH ;
HARVEY, TC .
PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (01) :88-95
[7]
A COMPARISON OF CF-252 AND PU-238, BE NEUTRON SOURCES FOR PARTIAL-BODY INVIVO ACTIVATION-ANALYSIS [J].
MORGAN, WD ;
VARTSKY, D ;
ELLIS, KJ ;
COHN, SH .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :413-424
[8]
A CLINICAL INSTRUMENT FOR MULTIELEMENT INVIVO ANALYSIS BY PROMPT, DELAYED AND CYCLIC NEUTRON-ACTIVATION USING CF-252 [J].
RYDE, SJS ;
MORGAN, WD ;
SIVYER, A ;
EVANS, CJ ;
DUTTON, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (10) :1257-1271
[9]
SCOTT MC, 1983, T AM NUCL SOC, V44, P33
[10]
INVIVO ELEMENTAL ANALYSIS IN OCCUPATIONAL-MEDICINE [J].
SCOTT, MC ;
CHETTLE, DR .
SCANDINAVIAN JOURNAL OF WORK ENVIRONMENT & HEALTH, 1986, 12 (02) :81-96