HIGH-SPEED BLACK BLOOD IMAGING OF VESSEL STENOSIS IN THE PRESENCE OF PULSATILE FLOW

被引:16
作者
CHIEN, D [1 ]
GOLDMANN, A [1 ]
EDELMAN, RR [1 ]
机构
[1] HARVARD UNIV,BETH ISRAEL HOSP,DEPT RADIOL,330 BROOKLINE AVE,BOSTON,MA 02215
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1992年 / 2卷 / 04期
关键词
BLOOD VESSELS; STENOSIS OR OBSTRUCTION; FLOW ARTIFACT; PHANTOMS; PULSE SEQUENCES; RAPID IMAGING; VASCULAR STUDIES;
D O I
10.1002/jmri.1880020413
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Stenosis phantoms were created to study the ability of "black blood" methods to image a vessel stenosis in the presence of pulsatile flow. Black blood images were acquired with a modified TurboFLASH (fast low-angle shot) method that eliminates flow signal by applying a set of prepulses before segmented data acquisition. With this high-speed approach, imaging can be completed within 16 seconds. This technique was compared with conventional spin-echo black blood, gradient-echo black blood, and gradient-echo bright blood methods. Loss of flow signal, which extended beyond the site of the stenosis, was seen on the gradient-echo bright blood images. The pattern of signal loss varied with the type of stenosis. Flow voids were achieved with spin-echo black blood imaging; however, substantial ghosting artifacts were seen. With gradient-echo black blood imaging, it was difficult to eliminate all flow signal, particularly for in-plane flow. The modified TurboFLASH method produced high-quality black blood images in a fraction of the time needed for spin-echo imaging. It showed no ghosting artifacts even in the presence of pulsatile flow.
引用
收藏
页码:437 / 441
页数:5
相关论文
共 15 条
[1]  
Masaryk TJ, Modic MT, Ruggieri PM, Et al., Three‐dimensional (volume) gradient‐echo imaging of the carotid bifurcation: preliminary clinical experience, Radiology, 171, pp. 801-806, (1989)
[2]  
Ross JS, Masaryk TJ, Modic MT, Et al., Intracranial aneu‐rysms: evaluation by MR angiography, AJNR, 11, pp. 449-456, (1990)
[3]  
Axel L, Blood flow effects in magnetic resonance imaging, AJR, 143, pp. 1157-1166, (1984)
[4]  
Nishimura DG, Macovski A, Pauly JM, Considerations of MR angiography by selective inversion recovery, Magn Re‐son Med, 7, pp. 472-484, (1988)
[5]  
Edelman RR, Chien D, Atkinson DJ, Sandstrom J, Fast time‐of‐flight magnetic resonance angiography with improved background suppression, Radiology, 179, pp. 867-870, (1991)
[6]  
Lin W, Haacke ME, Smith AS, Enhanced MRA imaging with Gd‐DTPA and vessel tracking (abstr), Book of ab‐stracts: Society of Magnetic Resonance in Medicine 1991, (1991)
[7]  
Saloner D, Hanson WA, Tsuruda JS, Van Tyen R, Anderson CM, Lee RE, Application of a connected‐voxel algorithm to MR angiographic data, JMRI, 1, pp. 423-430, (1991)
[8]  
Anderson CM, Saloner D, Tsuruda JS, Et al., Artifacts in maximum‐intensity‐projection display of MR angiograms, AJR, 154, pp. 623-629, (1990)
[9]  
Kaufman L, Crooks LE, Sheldon P, Et al., Evaluation of NMR imaging for detection and quantification of obstruction in vessels, Invest Radiol, 17, pp. 554-560, (1982)
[10]  
Laub GA, Kaiser WA, MR angiography with gradient motion rephasing, J Comput Assist Tomogr, 12, pp. 377-382, (1988)