CORONARY-ARTERY RUPTURE AND PSEUDOANEURYSM FORMATION RESULTING FROM PERCUTANEOUS CORONARY ANGIOSCOPY

被引:13
作者
WOLFF, MR [1 ]
RESAR, JR [1 ]
STUART, RS [1 ]
BRINKER, JA [1 ]
机构
[1] JOHNS HOPKINS UNIV HOSP,DIV CARDIOL,CMSC 501,600 N WOLFE ST,BALTIMORE,MD 21205
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1993年 / 28卷 / 01期
关键词
ANGIOSCOPY; CORONARY PSEUDOANEURYSM; OCCLUSION BALLOON;
D O I
10.1002/ccd.1810280110
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We describe a case in which coronary angioscopy was complicated by inability to deflate the device's occlusion balloon. Rapid over-inflation to rupture the balloon resulted in massive dissection of the artery, pseudoaneurysm formation and ultimately coronary bypass. While the cause of failure of balloon deflation remains obscure, deliberate over-inflation to cause rupture may be hazardous.
引用
收藏
页码:47 / 50
页数:4
相关论文
共 14 条
[1]   CARDIAC-TAMPONADE FOLLOWING PERFORATION OF THE LEFT ANTERIOR DESCENDING CORONARY SYSTEM DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - SUCCESSFUL TREATMENT BY PERICARDIAL DRAINAGE [J].
ALTMAN, F ;
YAZDANFAR, S ;
WERTHEIMER, J ;
GHOSH, S ;
KOTLER, M .
AMERICAN HEART JOURNAL, 1986, 111 (06) :1196-1197
[2]   ACUTE CORONARY EVENTS ASSOCIATED WITH PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
COWLEY, MJ ;
DORROS, G ;
KELSEY, SF ;
VANRADEN, M ;
DETRE, KM .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (12) :C12-C16
[3]   ANGIOSCOPY AFTER CORONARY EXCIMER LASER ANGIOPLASTY [J].
DIETHRICH, EB ;
HANAFY, HM ;
SANTIAGO, OJ ;
BAHADIR, I .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (02) :643-643
[4]   EXPERIMENTAL AND CLINICAL PERCUTANEOUS ANGIOSCOPY EXPERIENCE WITH DYNAMIC ANGIOPLASTY [J].
GEHANI, AA ;
BALL, SG ;
LATIF, AB ;
DAVIES, GA ;
REES, MR .
ANGIOLOGY, 1990, 41 (10) :809-816
[5]  
JOHNSON C, 1989, Journal of the American College of Cardiology, V13, p146A
[6]   TRANS-LUMINAL CORONARY ANGIOPLASTY COMPLICATED BY CORONARY-ARTERY PERFORATION [J].
KIMBIRIS, D ;
ISKANDRIAN, AS ;
GOEL, I ;
BEMIS, CE ;
GEHL, L ;
OWENS, J ;
SEGAL, BL .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1982, 8 (05) :481-485
[7]   ANGIOSCOPIC EVALUATION OF CORONARY-ARTERY THROMBI IN ACUTE CORONARY SYNDROMES [J].
MIZUNO, K ;
SATOMURA, K ;
MIYAMOTO, A ;
ARAKAWA, K ;
SHIBUYA, T ;
ARAI, T ;
KURITA, A ;
NAKAMURA, H ;
AMBROSE, JA .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (05) :287-291
[8]   PERCUTANEOUS ANGIOSCOPY DURING CORONARY ANGIOPLASTY USING A STEERABLE MICROANGIOSCOPE [J].
RAMEE, SR ;
WHITE, CJ ;
COLLINS, TJ ;
MESA, JE ;
MURGO, JP .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (01) :100-105
[9]   RUPTURE OF A PSEUDOANEURYSM OF A SAPHENOUS-VEIN CORONARY ARTERIAL BYPASS GRAFT PRESENTING WITH SUPERIOR CAVAL VENOUS OBSTRUCTION [J].
ROSIN, MD ;
RIDLEY, PD ;
MAXWELL, PH .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1989, 25 (01) :121-123
[10]   CORONARY ARTERIAL RUPTURE DURING CORONARY ANGIOPLASTY [J].
SAFFITZ, JE ;
ROSE, TE ;
OAKS, JB ;
ROBERTS, WC .
AMERICAN JOURNAL OF CARDIOLOGY, 1983, 51 (05) :902-904