ABSOLUTE MEASURES OF IMAGE QUALITY FOR THE SENS-A-RAY DIRECT DIGITAL INTRAORAL RADIOGRAPHY SYSTEM

被引:14
作者
WELANDER, U
MCDAVID, WD
MORNER, AC
TRONJE, G
TOKUOKA, O
FUCHIHATA, H
NELVIG, P
DOVE, SB
机构
[1] UNIV TEXAS,HLTH SCI CTR,DEPT DENT DIAGOST SCI,SAN ANTONIO,TX
[2] OSAKA UNIV,DEPT ORAL RADIOL,OSAKA,JAPAN
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS | 1995年 / 80卷 / 03期
关键词
D O I
10.1016/S1079-2104(05)80393-2
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
To study the noise characteristics of the Sens-A-Ray (Regam Medical Systems AB, Sundsvall, Sweden) system, 20 radiographs were obtained at each of three different exposure levels at 70 and 90 kVp with a homogeneous x-ray field. Exposures were measured with an ionization chamber. Noise power spectra were calculated over three areas within each radiograph, and ensemble averages were subsequently found from 60 data files at each exposure level. Noise equivalent quanta were calculated with the noise power spectra acid modulation transfer function data from previous studies. Finally, the detective quantum efficiency was calculated by dividing the noise equivalent quanta by The estimated incident photon fluence at the different exposures. The system has a maximum detective quantum efficiency of approximately 0.030 at 70 kVp and 0.025 at 90 kVp. A broad maximum exists at approximately 2 cycles/mm, indicating that the signal-to-noise ratio is most favorable at this spatial frequency.
引用
收藏
页码:345 / 350
页数:6
相关论文
共 16 条
[1]   A PHYSICAL IMAGE QUALITY EVALUATION OF A DIGITAL SPOT FLUOROGRAPHY SYSTEM [J].
COWEN, AR ;
WORKMAN, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (02) :325-342
[2]  
DOI K, 1982, 828187 HLTH HUM SERV
[3]  
MCDAVID WD, IN PRESS ORAL SURG O
[4]   AN EVALUATION OF THE SENS-A-RAY DIGITAL DENTAL IMAGING-SYSTEM [J].
MCDONNELL, D ;
PRICE, C .
DENTOMAXILLOFACIAL RADIOLOGY, 1993, 22 (03) :121-126
[5]   THERAPY IMAGING - A SIGNAL-TO-NOISE ANALYSIS OF METAL PLATE FILM DETECTORS [J].
MUNRO, P ;
RAWLINSON, JA ;
FENSTER, A .
MEDICAL PHYSICS, 1987, 14 (06) :975-984
[6]   SENS-A-RAY - A NEW SYSTEM FOR DIRECT DIGITAL INTRAORAL RADIOGRAPHY [J].
NELVIG, P ;
WING, K ;
WELANDER, U .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1992, 74 (06) :818-823
[7]   SIGNAL-TO-NOISE PROPERTIES OF MAMMOGRAPHIC FILM SCREEN SYSTEMS [J].
NISHIKAWA, RM ;
YAFFE, MJ .
MEDICAL PHYSICS, 1985, 12 (01) :32-39
[8]   FRACTAL DIMENSION FROM RADIOGRAPHS OF PERIDONTAL ALVEOLAR BONE - A POSSIBLE DIAGNOSTIC INDICATOR OF OSTEOPOROSIS [J].
RUTTIMANN, UE ;
WEBBER, RL ;
HAZELRIG, JB .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1992, 74 (01) :98-110
[9]   ABSOLUTE MEASURES OF PHYSICAL IMAGE QUALITY - MEASUREMENT AND APPLICATION TO RADIOGRAPHIC MAGNIFICATION [J].
SANDRIK, JM ;
WAGNER, RF .
MEDICAL PHYSICS, 1982, 9 (04) :540-549
[10]   NOISE IN D-SPEED AND E-SPEED RADIOGRAPHIC FILM [J].
SOUTHARD, TE ;
SOUTHARD, KA .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1993, 75 (06) :767-773